RayTek MI Miniature Infrared Sensor User Manual

Page of 95
Programming Guide
 
78 
MI 
Description 
Char  Format 
P  B  S  Legal values 
Factory 
default 
LCD 
Source: emissivity / 
setpoint for alarm output 
ES X 
*  * 
I=constant 
number 
(E=0.950) 
E=external 
analogous input 
FTC1 
D= E/XS digital 
selected FTC1-3 
I  
Presel. emissivity value 
EV  n.nnn 
 
*  0.100 - 1.100 
 
 
Valley hold time(4) 
F  nnn.n 
*  *  *  0.000 - 998.9 s 
(999 = infinite) 
000.0 s 
Average time 
G  nnn.n 
*  *  *  000.0 – 999 s 
000.0 s 
Top of mA/mV range 
H  nnn.n 
*  *  *  (1) 
(7) 
Sensor / head ambient 
nnn.n 
*  * 
  in current scale 
(°C/°F) 
 
 
Switch panel lock 
J  X 
 
*  L=locked 
U=unlocked 
unlocked  
Alarm output control 
K  N 
 
*  0=off 
1=on 
2=Target.; norm. 
open 
3=norm. closed 
4=Head; normal 
open 
5=norm. closed 
7=sensor / head 
ambient 
?  
Bottom of mA/mV range 
L  nnn.n 
*  *  *  (1) 
(8) 
Output voltage 
O  n.nnn 
 
*  0.000 – 5.000 
voltage in V 
6=controlled by 
unit  
6  
 
Output current 
O  nn.nn 
 
*  0.00 – 20.00 
current in mA 
21=over range 
60=controlled by 
unit 
60 L 
Peak hold time (4) 
P  nnn.n 
*  *  *  000.0 998.9 s 
(999 = infinite) 
000.0 s 
Power/AD value 
Q  nnnn 
*  *