5.
Operation
Meaning of indication:
Operator activity:
yes
no
no
Acknowledge indication by
pressing 
 once briefly.
Initialization phase 1
Initialization phase 2
Software version
Selected display language
Sectioning motor speed
level  ‘slow’ / (fast)
---
---
---
Select language via
keys
Select speed level via
keys
5.3.3 Display indications when switching on
Standard indication ‘locked’ is displayed:
(Display background dark)
To unlock: press 
 until the display back-
ground lights up:
yes
Temperature prior to
switching off < 0 °C
If this indication is displayed: Do not switch on
the instrument!
Prior to switching on, both microtome and
accessories must be completely dry!
Remove microtome/accessories, disinfect if
applicable, then dry thoroughly and reinstall.
(--> See chapters 7.1 - 7.5).
*) Indication in instruments with specimen cooling. Instruments without specimen cooling: only CT (centered) is displayed.
Temperature upon
switching on 
≥ 0 °C
8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8
8 8 8 8 8 8 8 8 8 8 8 8 8 8 8 8
               
               
C M 3 0 5 0 S
V s . X X . X X
E N G L I S H
C U T T I N G
M O T O R
S L O W
P O W E R
F A I L U R E
1 2 : 0 0
C T . . . . ° C
O T . . . . ° C
  H O M E
❄ ❄
C T . .
. . ° C
O T . .
. . °  C  
* )
D R Y
M I C R O T O M E
C T . . . . ° C
O T . . . . °  C  
* )
C T . .
. . ° C
O T . .
. . °  C  
* )