Leica CM 3050 S – Cryostat
5.
Operation
Press this key to alternate between the
two functions.
two functions.
illuminated:
Trimming mode is active
Trimming mode is active
off:
Sectioning mode is active.
Sectioning mode is active.
Selecting trimming thickness
•
Activate trimming mode
•
Select trimming thickness (= sectioning
thickness at which trimming is carried out)
by pressing
thickness at which trimming is carried out)
by pressing
or
.
The selected thickness is displayed in the cor-
responding LED
responding LED
:
Motorized trimming
•
Select ‘Continuous stroke’ in functional area
A.
A.
•
Activate trimming by pressing RUN/STOP
and RUN/ENABLE (located in functional
area C) simultaneously or by depressing the
footswitch.
and RUN/ENABLE (located in functional
area C) simultaneously or by depressing the
footswitch.
•
To stop trimming press RUN/STOP or RUN/
ENABLE or depress the footswitch again.
ENABLE or depress the footswitch again.
Trimming thickness can be changed while trim-
ming is in motion by pressing
ming is in motion by pressing
or
.
5.4.7 Setting the section thickness
•
Activate sectioning mode (
off).
•
Press
or
to select section thickness.
Pressing one of these keys once makes the
display in control panel 1 switch to section
thickness indication (thickness value not yet
being altered). Pressing
display in control panel 1 switch to section
thickness indication (thickness value not yet
being altered). Pressing
or
further
starts to change the actual section thick-
ness setting.
ness setting.
Selectable section thickness settings:
0
-
2 μm
in 0.5-μm
steps,
2
-
10 μm
in 1-μm
steps,
10
-
20 μm
in 2-μm
steps,
20
-
60 μm
in 5-μm
steps,
60
- 100 μm
in 10-μm
steps,
100
- 300 μm
in 50-μm
steps.
5.4.6 Trimming and sectioning functions
Functional area B - all models
Switching between trimming
and sectioning
and sectioning
C T . . . . ° C
O T . . . . ° C
μ 0 1 0 . 0