Leica CM3050 S User Manual

Page of 66
54
Instruction Manual V1.2 - 11/2006
7
Disinfection, cleaning and maintenance
7.3
Disinfection with conventional
Switch the instrument off.
Carefully remove knife / blade as well as spec-
imens and accessories from the chamber.
Remove section waste from the chamber
and dispose of it.
Remove sliding window (see chapter 7.6 ‘Ex-
changing the fluorescent light lamp’).
Remove microtome (see chapter 7.5.1
‘Removing the microtome’).
Allow the microtome to thaw, disinfect and
allow to dry thoroughly.
Drying the microtome:
• Use a hairdryer: dry and before
reinstalling into the chamber
allow the microtome to cool
down to room temperature.
• Allow to dry at room temperature
(at least over night - longer,
if necessary).
Disinfect and dry disassembled parts
(accessories and tools).
Disinfect chamber walls with commercial
alcohol disinfectant.
Place an adequate container underneath the
drain tube at the right side of the instrument.
Pull stopper out of chamber bottom and
drain off disinfecting solution into the con-
tainer.
Put stopper back into drain in cryochamber.
disinfectants
Allow cryochamber to dry thoroughly.
Reinstall microtome (see chapter 7.5.3).
Only when they are completely dry, reinsert
the disinfected accessories / tools into the
cryochamber.
Reinsert the sliding window.
Switch the instrument back on.
Once a suitable chamber temperature has
been reached, return the specimens to the
cryochamber.
7.4
Cleaning
When necessary, the instrument surfaces
can be cleaned with mild commercial house-
hold cleaners.
Surfaces which are potentially contaminated
(infectious germs) should always be cleaned
with alcohol disinfectants!