Leica CM3050 S User Manual

Page of 66
55
Leica CM 3050 S – Cryostat
7.5
Removing / reinstalling the microtome
7.5.1 Removing the microtome
When removing the microtome
make sure to observe the following
safety instructions:
• ‘Removing / reinstalling
the microtome’
(Chapter 2.3.11)
• ‘Defrosting / handling frozen
tissue’
(Chapter 2.3.7)
• ‘Infectious / radioactive material’
(Chapter 2.3.9)
• ‘Disinfection and cleaning’
(Chapter 2.3.10)
• Prior to removing the microtome, carry out
preparation steps as per chapter 7.3 (switch
instrument off, remove accessories etc.)
7.5.2 Disassembling the specimen head
Use Allen key size 5 to remove screw (2)
from the center of the specimen head (1).
Do not twist the refrigerating tube (see
chapter  2.3.11 ‘Removing / reinstalling the
microtome’).
Hold hook (4) (part of standard delivery) in
one of your hands and with your other hand
rest refrigerating tube (3) on hook.
Insert hook (4) with refrigerating tube (3) into
the opening above mark (
▼).
7.
Disinfection, cleaning and maintenance
(instruments w/specimen cooling only)