11 
simply multiplied values, thus the same effect can be added later on in image editing 
software. Therefore, changing the value from 100 is not recommended. 
3.3.3
 
Exposure Mode 
In the Exposure Mode box, select either Automatic or Manual
 
Automatic  (AE):  The  software  measures  the  optimal  exposure  time 
automatically.  Change  Spot  and  Exposure  Adjust  if  needed  (see  below).  To 
lock  the  exposure  time,  click  AE  lock  button  (Figure  15,  upper  toolbar, 
rightmost button).  
o
 
Spot defines the size of an area from which exposure time is measured. 
Choose 30 %, 1 % or 0.1 % of picture area. To move the spot, there is 
a  button  in  the  upper  toolbar  (Figure  15,  upper  toolbar,  third  button 
from right). Click the button and then click live view image in a place 
where you want the exposure area to be. Re-centering the area is done 
with the button on the right-hand side of the Spot button. 
o
 
Exposure  Adjust  is  used  for  adjusting  the  automatic  exposure  time. 
Use  values  from  0  to  2  to  brighten  (longer  exposure)  the  image  and 
values from -2 to 0 to darken (shorter exposure) the image.  
 
Manual:  You  can  adjust  the  exposure  time  manually  using  the  Exposure 
Time
 slider. 
3.3.4
 
White balance 
White balance is the adjustment of relative intensities in red, green, and blue channels. 
White  balance  adjustment  is  needed  because  the  halogen  lamp  in  the  microscope 
changes  color  with  the  change  of  intensity.  The  dimmer  the  light,  the  more  red  it 
becomes. 
Pressing the button marked with OWB will automatically adjust the white point of the 
image. If manual adjustment is needed, press the WB-button and make a selection in 
the  image  that  represents  white  color.  The  image  is  adjusted  accordingly.  The  best 
white balance is achieved by pressing the OWB-button while the sample is out of the 
view (the whole camera view area is white) so it does not disturb the measurement of 
the color of the light.  
3.3.5
 
Levels adjustment 
You  can  adjust  the  brightness,  contrast,  and  gamma  correction  of  the  image  in  the 
levels  adjustment  dialog  (Main  Menu/BF-button/Settings/Level  Adjustment  tab). 
Start  the  adjustment  by  clicking  the  levels  button  (marked  with  red  square  in  Figure 
16). The window contains an image histogram, which shows the intensity distribution 
of  the  image.  Wide  histogram  indicates  that  good  dynamic  range  of  the  camera  was 
used. A peak at the right end means that there are oversaturated pixels in the image. 
You should avoid over saturation by shortening the exposure time.  
The  Channel  selection  lets  you  adjust  one  color  at  a  time.  In  the  RGB  setting,  your 
adjustments  will  affect  all  colors.  Adjust  Shadow  and  Highlight  for  brightness  and 
contrast,  and  Gamma  for  gamma  correction.  Note  that  Gamma  is  a  non-linear 
change and has to be mentioned in the legend when the image is published.