Leica B523 User Manual

Page of 18
 
 
 
 
Figure 9. Light pointer controls 
2.8
 
Koehler illumination 
Koehler illumination needs to be adjusted in the  beginning of every imaging session 
for optimal illumination. Proceed as follows: 
Instruction 
 
View in eyepiece 
1.  Use  10x  objective  and 
front  lens.  Focus  on  the 
sample. 
2.  Close  the  luminous-field 
diaphragm  by  turning  the 
adjustment wheel clockwise. 
 
3.  Move  the  condenser  up or 
down  so  that  the  diaphragm 
edges  are  in  focus.  The 
microscope  is  designed  so 
that in the imaging path, field 
diaphragm  should  be  in 
conjugate  plane  with  the 
specimen  plane  (in  focus  at 
the  same  time).  You  adjust 
this by adjusting the height of 
the condenser.