Maintenance 
Instruction 3015-4342 
37
 
5 Maintenance 
5.1 Diagnostics Menu 
The Diagnostics menu provides information about the instrument’s operation that will be useful to service 
technicians for the purpose of troubleshooting various fault conditions. 
 
To display the Diagnostics menu, first press both the ENT and ESC buttons on either the instrument’s 
front panel or Smart probe at the same time to display the Main menu. Next, use the Up and Down 
buttons to highlight “DIAGNOSTICS” and then press ENT to display the Diagnostics menu. 
 
Main Menu:  
 
SYSTEM 
 LOG 
 SETUPS 
 CAL 
DATA 
 DIAGNOSTICS 
 
 
 
 
 
 
 
5.1.1 Sensor Data 
The Sensor Data screen provides information about the sensor’s operating parameters and its surrounding 
environment. When contacting the Bacharach Service Department, the operator may be asked to provide 
this information to help in the troubleshooting process. 
 
To display the Sensor Data screen, first display the Diagnostics menu as described in Section 5.1. Next, 
use the Up and Down buttons to highlight “SENSOR DATA” and then press ENT
 
Sensor Data Screen Information: 
LR
 .......................... Measured Leak Rate 
FAULT
 ................... Current Numerical Fault Code (refer to Sectio5.1.2 Last Fault
OPEN/CLOSED
  .... Measure / Search Flow Rate Valve Position 
DET
 ....................... Detector Voltage 
N
 ............................ Detector’s Output Noise Factor (with no gas detected) 
Z
  ............................ Stored Zero Voltage of Detector 
DET
 ....................... Detector Temperature 
SPPM
..................... PPM Corrected to Standard Pressure and Temperature 
BOX
 ....................... Enclosure Temperature 
M
 ........................... Manifold Pressure 
AMB
....................... Ambient Pressure 
MPM
....................... Moles Per Minute 
AU
.......................... Average Optical Absorbency Unit of Gas being detected (0 = no gas) 
PPM
........................ Parts Per Million value of the Current Detected Gas Level 
UML
........................ Absolute Concentration of Gas in Micro Moles per Liter 
 
 
 
 
 
 
Diagnostics Menu:  
 
SENSOR DATA 
 LAST 
FAULT 
 IR 
EMITTER 
 PRESSURE