Bacharach H25-IR Manual De Usuario

Descargar
Página de 60
Maintenance 
H25-IR 
 
38
 
Instruction 3015-4342 
5.1.2 Last Fault 
The Last Fault screen provides a text description of all current faults that have been detected by the 
instrument. Once the cause of the fault(s) has been cleared, its description is removed from the screen. The 
name of the last fault, however, is retained on the screen. This is useful in determining the cause of an 
intermittent type fault. 
To display the Last Fault screen, first display the Diagnostics menu as described in Section 5.1. Next, use 
the Up and Down buttons to highlight “LAST FAULT” and then press ENT.  If necessary, use the Up and 
Down buttons to scroll through the fault list. A list of possible fault messages is shown below. 
The numbers in brackets “< >” are the fault numerical codes that appear on the Diagnostics Sensor Data 
screen (refer to Section 5.1.1).  
Fault Codes are Additive:
 The fault codes 
appearing on the Sensor Data, Diagnostics 
screen (refer to Section 5.1.1) are additive. For 
example: A fault code of <0810> indicates that 
both a Search Flow Fault <0800> and a Loop 
Open error <0010> have occurred. 
 
Box Temp Error <0001>:
 Enclosure temperature 
is high. Check that the enclosure is not being 
subjected to extreme temperatures. Use the 
Sensor Data screen to check the box temperature  
(refer to Section 5.1.1).  
 
Bench Temp Error <0002>:
 Optical bench 
temperature is high. Check that the enclosure is 
not being subjected to extreme temperatures. 
Use the Sensor Data screen to check the box 
temperature (refer to Section 5.1.1).  
 
Loop Open <0010>:
 The 4−20 mA current loop is 
open, or there is a high resistance in the circuit. 
Check the wiring to the load/monitoring circuit. 
If this feature is not being used, a 100 ohm 
resistor must be connected to the 4−20 mA 
output connector (refer to Section 2.8). 
 
Internal CAL Error <0020>:
 An internal 
calibration failure has occurred. Repeat internal 
calibration procedure per Section 3.2. 
 
External CAL Error <0040>:
 An external calibra-
tion failure has occurred. Repeat external cali-
bration procedure per Section 3.3. 
 
Digipot Range Error <0200>:
 The digital 
potentiometer circuit is out of its normal 
operating range. Call Bacharach Service with 
this information for further instructions. 
 
Leak Pressure Fault <0100>:
 The pressure of the 
internal leak source has dropped below its 
minimum value. Call Bacharach Service with 
this information for further instructions. 
 
 
A/D Failure <0400>:
 A failure has occurred in the 
instrument’s analog-to-digital circuitry. Call 
Bacharach Service with this information for 
further instructions. 
 
Search Flow Fault <0800>:
 This error occurs in 
the search mode when the gas flow through the 
detector has significantly changed. Check that 
the probe connector is plugged into the instru-
ment, the probe hose is not kinked, and the filter 
in the probe tip is not clogged. Also check for a 
blocked internal filter or failed pump. 
 
Measure Flow Fault <1000>:
 This error occurs in 
the measure mode when the gas flow through the 
detector has significantly changed. Check that 
the probe connector is plugged into the instru-
ment, the probe hose is not kinked, and the filter 
in the probe tip is not clogged. Also check for a 
blocked internal filter or failed pump. 
 
No Flow <2000>:
 Gas flow through the detector 
has stopped. Check that the probe hose is not 
kinked, and the filter in the probe tip is not 
clogged. Also check for a blocked internal filter or 
failed pump. 
 
Zero Volt Error <4000>:
 The detector’s output 
voltage may be out of tolerance. Use the Sensor 
Data screen to check the detector voltage (refer 
to Section 5.1.1). Call Bacharach Service with 
this information for further instructions. 
 
A/D Clipping Error <8000>:
 The detector voltage 
may be out of tolerance. Use the Sensor Data 
screen to check the detector voltage (refer to 
Section 5.1.1). Call Bacharach Service with this 
information for further instructions.