Configure process measurement
Topics covered in this chapter:
4.1
Configure mass flow measurement
The mass flow measurement parameters control how mass flow is measured and reported.
The mass flow measurement parameters include:
Mass Flow Measurement Unit
Flow Damping
Mass Flow Cutoff
4.1.1
Configure Mass Flow Measurement Unit
Display (standard)
OFF-LINE MAINT
 > OFF-LINE CONFG > UNITS > MASS
Chinese-language
display
Offline Maintain
 > Configuration > Units > Mass Flow Rate
ProLink II
ProLink
 > Configuration > Flow > Mass Flow Unit
ProLink III
Device Tools
 > Configuration > Process Measurement > Flow
Field Communicator Configure > Manual Setup > Measurements > Flow > Mass Flow Unit
Overview
Mass Flow Measurement Unit
 specifies the unit of measure that will be used for the mass flow
rate. The unit used for mass total and mass inventory is derived from this unit.
Procedure
Set Mass Flow Measurement Unit to the unit you want to use.
The default setting for Mass Flow Measurement Unit is g/sec (grams per second).
Configure process measurement
Configuration and Use Manual 