Toshiba LF414 Manual De Instrucciónes

Descargar
Página de 65
 
 
−  36  − 
 
6F8A0870
 
Notes on wiring 
 
CAUTION
 
(1)  Select the cable runs away from electrical equipment (motors, transformers, or radio 
transmitters) which causes electromagnetic or electrostatic interference
(2)  Deterioration of flowmeter circuit insulation occurs if the converter interior or cable ends get wet or 
humidified. This in turn causes malfunction of flowmeter or noise problemsAvoid a rainy day if 
the flowmeter is to be installed outdoors
. Even indoors, prevent water from splashing over the 
flowmeter. Try to finish the wiring as quickly as possible 
(3)  The converter has an arrestor installed inside. Therefore, do not conduct a withstand voltage test 
for the converter. To check the insulation of the converter, use a voltage of 250Vdc or less
(4)  After wiring, be sure to install the terminal block protection cover. 
(5)  Because the excitation cable and flow rate signal cable transmit very delicate signals, pass each of 
them separately through a thick steel conduit tube, keep them away from the large current wiring as 
far as possible, and do not install them in parallel. 
 
 
5.1 Cables 
 
Use the kind of cables shown in Table 5.1 to wire the converter. 
Table 5.1 Installation Cables 
Name Cable 
name 
Nominal 
cross-section
al area 
Finished 
outer 
diameter 
Description 
Power cable 
3-core vinyl sheathed cable or 2-core vinyl 
sheathed cable 
2 mm² 
11~13mm 
CVV JIS C 3401 or 
equivalent 
Output 
signal cable 
The number of conductors the cable contains differs depending on the 
specification of the output signal cable. 
Use a shielded cable of finished outer diameter 11 to 13mm and 
nominal cross-sectional area 1.25mm
2
CVV-S JIS -258-C or 
equivalent 
 
Flow rate 
signal cable 
2-core shielded chloroprene cabtyre cable 
( Rubber covered cable ) 
0.75 mm
2
 11~13mm 
2PNCT-S 
JIS C 3327 or 
equivalent
 
 
Excitation 
cable 
3-core chloroprene cabtyre cable 
( Rubber covered cable ) 
2 mm
1.25 m
2
 
11~13mm 
2PNCT 
JIS C 3327 or 
equivalent
 
 
Separate 
Separate 
Separate