−  38  − 
 
6F8A0870
 
5.4 Wiring 
 
5.4.1 Terminal Treatment of Cables 
 
Follow the procedures below to treat the terminals (at the converter side) of various cables and install the 
cables to the terminal block. Use appropriate cables based on the description in Section 5.1 "Cables." Crimp 
round type insulated crimp-type terminal to the end of the cables. 
 
(1) Power cable, current output cable, digital I/O cables 
 
The necessary cables should be ordered from the person responsible for the installation. Strip the sheath of 
each conductor as shown in Figure 5.1 and attach a crimping terminal with insulated sleeve to it. The size of 
the crimping terminal is as follows:   
M4:    Integral type LF600F and LF610F 
M3.5: Integral type LF620F, Separate type LF602F, LF612F and LF622F 
 
•  Connect the power cable to terminal blocks L1 and L2. 
•  Connect the current output cable to terminal blocks + and -. 
•  Connect the digital I/O cable to terminal blocks D1, D01, D02 and COM, as required. 
 
                     
   
 
       Figure 5.1  Terminal Treatment of Power Cable, Current Output Cable  
and Digital I/O cable 
 
(2) Excitation cable 
 
Strip the sheath from the end of each conductor as shown in Figure 5.2, attach an M3.5 crimping terminal 
with insulated sleeve, and connect it to the terminal blocks X and Y. Connect the red conductor to terminal 
block E. 
                       
 
 
Figure 5.2    Terminal Treatment of Excitation Cable 
 
Crimping terminal 
 25~45mm 
M3.5 crimping terminal 
X Black 
E Red 
Y White 
 25~45mm 
M4: LF600F and LF610F 
M3.5: LF602F, LF612F, LF620F and LF622F
Separate