GE PT878GC Manual De Usuario

Descargar
Página de 308
April 2004
9-38
Servicing the PT878GC
Testing the Watchdog Timer Circuit
The PT878GC includes a watchdog timer circuit. If a software 
error causes the meter to stop responding, this circuit 
automatically resets the meter. A properly functioning PT878GC 
restarts if you run the Watchdog Test.
To start the Watchdog Test, scroll to the Watchdog Test option in 
the Test Menu and press 
[ENTER]
. The screen appears similar to 
Figure 9-23 below.
Figure 9-23:  The Watchdog Test Screen
Press 
[F2]
 (No) to cancel the test and return to the Menu screen, 
or press 
[F3]
 (Yes) to start the test. The PT878GC should go 
blank for a few seconds, and then restart. If it does not follow this 
sequence, consult GE Panametrics.
Dew Point, °C
Signal, dB
-52.0
32
Pressure, PSIg
55.00
10.0
Run Time, Hours
Site   Program   Meter  Logging   
Service
Units
Reports
Reference
MIS Probe
Factory Defaults
Flash Update
0
6
12
Factory Calibration
Test                   
X
No
Watchdog Test
Test Screen
Test Keys
Yes
** WARNING **
This test should automatically
restart your meter!
Press ‘Yes’ to start the test.
Press ‘No’ to cancel the test.
Watchdog Test