Servicing the PT878GC
9-39
April 2004
Setting Impulse Response
The Impulse Response option enables you to force the meter to 
transmit in one direction only, without changing the AGC setting. 
You can then diagnose problems with transducer, pipe, or fluid 
configurations. To enter the option, scroll down to the Impulse 
Response entry in the Service Menu and press 
[ENTER]
. The 
screen appears similar to Figure 9-24 below.
Figure 9-24:  The Impulse Response Option
1. The first prompt asks in which direction you wish the meter to 
transmit, upstream or downstream.
a. Press 
[ENTER]
 to open the drop-down list.
b. Use the  [
T
] or [
S
] arrow keys to scroll to the desired 
direction.
c. Press 
[ENTER]
 to confirm your selection.
Transmit
  
 
Stop
Exit
Impulse Response
 
 
 
Direction
 
Upstream
Percent Gain
           
 20
%
Stopped.