NI 4070/4072 Calibration Procedure
38
ni.com
To verify the capacitance measurements of the NI 4072, complete the 
following steps:
1.
Disconnect any fixtures or cables from the NI 4072.
2.
Call 
niDMM_reset
 to reset the NI 4070/4072 to a known state.
3.
Call 
niDMM_ConfigureMeasurement
 with the following 
parameters:
Function 
NIDMM_VAL_CAPACITANCE
Range 
300e-12
Resolution 
50e-15
4.
Set the number of averages of the NI 4072 to 
20
 by calling 
niDMM_SetAttributeViInt32
 with the following parameters:
Attribute_ID 
NIDMM_ATTR_LC_NUMBER_MEAS_TO_AVERAGE
Attribute_Value 
20
Note
This measurement corresponds to a 0 pF capacitance.
5.
Call 
niDMM_Read
. Verify that this measurement falls between the 
tolerances listed in Table 22.
6.
Plug in the insulated banana plug shorting bar across the HI and LO 
banana plug connectors of the NI 4072.
7.
Call 
niDMM_ConfigureMeasurement
 with the following 
parameters:
Function 
NIDMM_VAL_INDUCTANCE
Range 
10e-6
Resolution 
1e-9
8.
Set the number of averages of the NI 4072 to 
40
 by calling 
niDMM_SetAttributeViInt32
 with the following parameters:
Attribute_ID = 
NIDMM_ATTR_LC_NUMBER_MEAS_TO_AVERAGE
Attribute_Value = 
40
Note
This measurement corresponds to a 0
μH inductance.
9.
Call 
niDMM_Read
. Verify that this measurement falls between the 
tolerances listed in Table 22.
10. Remove the shorting bar and plug the 270 pF verification capacitor 
into the HI and LO banana plug connectors of the NI 4072. Remember 
to wait 30 seconds for the temperature to stabilize before performing 
the next step.