NI 4070/4072 Calibration Procedure
11. Call 
niDMM_ConfigureMeasurement
 with the following 
parameters:
Function 
NIDMM_VAL_CAPACITANCE
Range 
300e-12
Resolution 
50e-15
12. Set the number of averages of the NI 4072 to 
20
 by calling 
niDMM_SetAttributeViInt32
 with the following parameters:
Attribute_ID = 
NIDMM_ATTR_LC_NUMBER_MEAS_TO_AVERAGE
Attribute_Value = 
20
13. Call 
niDMM_Read
. Verify that this measurement falls between the 
tolerances listed in Table 22.
Note
If you use capacitance verification values that differ from the values listed in 
Table 10, verify that each measurement falls between the tolerances listed in Table 22. 
The tolerances shown in Table 22 correspond to the NI 4072 accuracy specifications.
14. Remove the 270 pF verification capacitor, and plug the 1 nF 
verification capacitor into the HI and LO banana plug connectors of the 
NI 4072.
15. Call 
niDMM_ConfigureMeasurement
 with the following 
parameters:
Function 
NIDMM_VAL_CAPACITANCE
Range 
1e-9
Resolution 
100e-15
16. Set the number of averages of the NI 4072 to 
20
 by calling 
niDMM_SetAttributeViInt32
 with the following parameters:
Attribute_ID = 
NIDMM_ATTR_LC_NUMBER_MEAS_TO_AVERAGE
Attribute_Value = 
20
Note
If you use capacitance verification values that differ from the values listed in 
Table 10, verify that each measurement falls between the tolerances listed in Table 22. 
The tolerances shown in Table 22 correspond to the NI 4072 accuracy specifications.
17. Call 
niDMM_Read
. Verify that this measurement falls between the 
tolerances listed in Table 22.
18. Call 
niDMM_ConfigureMeasurement
 with the following 
parameters:
Function 
NIDMM_VAL_CAPACITANCE
Range 
10e-9
Resolution 
1e-12