NI 4070/4072 Calibration Procedure
11. Call
niDMM_ConfigureMeasurement
with the following
parameters:
•
Function =
NIDMM_VAL_CAPACITANCE
•
Range =
300e-12
•
Resolution =
50e-15
12. Set the number of averages of the NI 4072 to
20
by calling
niDMM_SetAttributeViInt32
with the following parameters:
•
Attribute_ID =
NIDMM_ATTR_LC_NUMBER_MEAS_TO_AVERAGE
•
Attribute_Value =
20
13. Call
niDMM_Read
. Verify that this measurement falls between the
tolerances listed in Table 22.
Note
If you use capacitance verification values that differ from the values listed in
Table 10, verify that each measurement falls between the tolerances listed in Table 22.
The tolerances shown in Table 22 correspond to the NI 4072 accuracy specifications.
The tolerances shown in Table 22 correspond to the NI 4072 accuracy specifications.
14. Remove the 270 pF verification capacitor, and plug the 1 nF
verification capacitor into the HI and LO banana plug connectors of the
NI 4072.
NI 4072.
15. Call
niDMM_ConfigureMeasurement
with the following
parameters:
•
Function =
NIDMM_VAL_CAPACITANCE
•
Range =
1e-9
•
Resolution =
100e-15
16. Set the number of averages of the NI 4072 to
20
by calling
niDMM_SetAttributeViInt32
with the following parameters:
•
Attribute_ID =
NIDMM_ATTR_LC_NUMBER_MEAS_TO_AVERAGE
•
Attribute_Value =
20
Note
If you use capacitance verification values that differ from the values listed in
Table 10, verify that each measurement falls between the tolerances listed in Table 22.
The tolerances shown in Table 22 correspond to the NI 4072 accuracy specifications.
The tolerances shown in Table 22 correspond to the NI 4072 accuracy specifications.
17. Call
niDMM_Read
. Verify that this measurement falls between the
tolerances listed in Table 22.
18. Call
niDMM_ConfigureMeasurement
with the following
parameters:
•
Function =
NIDMM_VAL_CAPACITANCE
•
Range =
10e-9
•
Resolution =
1e-12