Furuno M51A Manual De Usuario

Descargar
Página de 22
DLM 600W Series  
Isolated Remote Analog Interface Operation 
M51A Option 
2-7 
CURRENT RESISTANCE PROGRAMMING OUTPUT: Pin-22 provides a 1mA current source 
which would be connected to Pin-10, CURRENT PROGRAMMING INPUT, with a 0-5kΩ 
external resistor connected between Pin-22 to Pin-12, to program the output current from 
0-100% of full scale output.  The INTFC SETUP switch Position-2, I, must be set to OFF (down) 
to select 0-5VDC input range.  Circuit is SELV, and has electrical isolation from the output of the 
unit. 
Auxiliary Sources 
AUXILIARY 5VDC OUTPUT: Pin-15, 5VDC source output for use with logic and programming 
circuits. Source capability is adequate to provide for full scale programming of output voltage, 
current, or OVP when the user connects it to the appropriate programming input(s).  Source is 
referenced to Pin-6.  Maximum output current is 15mA.  Circuit is SELV, and has electrical 
isolation from the output of the unit.  
AUXILIARY 5VDC RETURN:  Pin-6, return of 5VDC AUXILIARY source output.  Circuit is 
SELV, and has electrical isolation from the output of the unit. 
1.25VDC REFERENCE FOR 4-20mA VOLTAGE PROGRAMMING: Pin-20, provides a 
1.25VDC reference utilized by internal circuits associated with 4-20mA programming of output 
voltage. 
1.25VDC REFERENCE FOR 4-20mA CURRENT PROGRAMMING: Pin-11, provides a 
1.25VDC reference utilized by internal circuits associated with 4-20mA programming of output 
current. 
2.4  Remote Programming Configuration 
Table 2–3 presents the options for remote control operation.  Pin numbers refer to the 
ISOLATED ANALOG INTERFACE connector.  Switch position numbers refer to the INTFC 
SETUP switch on the rear panel. 
 
Mode of 
Operation 
ANALOG-
CONTROL 
Signal, Pin-1 
LCK-OUT, INTFC 
Setup Switch 
Position-7 
Power-Up 
State 
Local/Remote 
Toggling 
Remote Only 
High Level 
ON (Up) 
Remote 
No 
Local Only 
Low or Open 
Off (Down) 
Local 
No 
Local/Remote 
High Level 
OFF (Down) 
Remote 
Yes 
Table 2–3.  Remote Programming Configuration Options