Isolated Remote Analog Interface Operation 
DLM 600W Series  
2-8 
M51A Option 
2.4.1 
2.4.2 
2.4.3 
Voltage Source Programming of Output Voltage 
 
Setting up for voltage source programming of the output voltage is as follows: 
 
1. 
Set Position-1, V, of the INTFC SETUP switch to OFF (down) for 0-5VDC programming 
range. 
 
2. 
Set Position-1, V, of the INTFC SETUP switch ON (up) for 0-10VDC programming 
range. 
 
3. 
Connect the external programming voltage source to the ISOLATED ANALOG 
INTERFACE connector, with positive to Pin-9 and negative to Pin-12. 
 
4. 
Program the other parameters to the desired limit values: CURRENT PROGRAMMING 
INPUT, Pin-10, and the OVP PROGRAMMING INPUT, Pin-3, with respect to pin-12. 
 
5. 
Connect Pin-1, ANALOG-CONTROL, of the ISOLATED ANALOG INTERFACE 
connector to Pin-15. 
 
Voltage Source Programming of Output Current 
 
Setting up for voltage source programming of the output current is as follows: 
 
1. 
Set Position-2, I, of the INTFC SETUP switch to OFF (down) for 0-5VDC programming 
range. 
 
2. 
Set Position-2, I, of the INTFC SETUP switch ON (up) for 0-10VDC programming range. 
 
3. 
Connect the external programming voltage source to the ISOLATED ANALOG 
INTERFACE connector, with positive to Pin-10 and negative to Pin-12. 
 
4. 
Program the other parameters to the desired limit values: VOLTAGE PROGRAMMING 
INPUT, Pin-10, and the OVP PROGRAMMING INPUT, Pin-3, with respect to pin-12. 
 
5. 
Connect Pin-1, ANALOG-CONTROL, of the ISOLATED ANALOG INTERFACE 
connector to Pin-15. 
 
Voltage 
Source 
Programming of OVP 
 
Setting up for voltage source programming of OVP is as follows: 
 
1. 
Set Position-3, OVP, of the INTFC SETUP switch to OFF (down) for 0-5VDC 
programming range. 
2. 
Set Position-3, OVP, of the INTFC SETUP switch ON (up) for 0-10VDC programming 
range. 
 
3. 
Connect the external programming voltage source to the ISOLATED ANALOG 
INTERFACE connector, with positive to Pin-3 and negative to Pin-12.