Trilogy200 clinical manual
46
Low Battery Alarm
 
The Low Battery alarm occurs when the last battery available is 
low or nearly depleted. This alarm occurs in two stages. When 
approximately 20 minutes of battery run time remains, a medium 
priority alarm is generated, and the device continues to operate. If 
no action is taken and the battery continues to deplete, the alarm 
escalates to a high priority alarm when approximately 10 minutes of 
battery run time remains. 
High Temperature Alarm 
 
This alarm occurs when the estimated patient airstream temperature 
or the ventilator internal temperature is too high. The alarm occurs in 
several stages. The ventilator continues to operate. Internal fans are 
started when the medium priority alarm is generated. If the condition 
causing the high temperature is not corrected and the temperature 
continues to rise, the alarm will escalate to the high priority alarm. 
Replace Detachable Battery Alarm
 
The Replace Detachable Battery alarm occurs when the detachable 
battery is nearing the end of its useful life or a failure in the 
detachable battery that prevents it from charging or discharging has 
been detected.
The alarm occurs in several stages, from low to high priority. The 
device continues to operate when the alarm is the low priority 
alarm. If the alarm is reset without removing the battery, the alarm 
will be regenerated once every hour until the detachable battery is 
removed. The device continues to operate, the detachable battery 
is not used, and the power source is switched to the next available 
power source if the alarm is the high priority alarm.
Ventilator Service Recommended Alarm
 
This is a medium priority alarm. It occurs when the device has 
detected an error, but the error will not affect device performance 
or safety. The device continues to operate. Therapy and safety are not 
compromised. 
 WARNING 
Immediately seek an alternate 
power source when the “Low 
Battery” message appears.  
Complete power failure and loss 
of therapy is imminent.
 WARNING
The ventilator has a two-stage 
low battery alarm. The medium 
priority alarm indicates that 
approximately 20 minutes of 
operation remain, and the high 
priority alarm indicates that less 
than 10 minutes of operation 
remain. Actual run time may be 
more or less than this and varies 
with battery age, environmental 
conditions, and therapy.