Trilogy200 clinical manual
48
Check External Battery Info Message 
This info message occurs when a bad connection exists to the 
external battery or the external battery failed. The device continues 
to operate using power from the detachable battery, if available, or 
the internal battery.   
Battery Depleted Info Message 
 
This info message occurs when the affected battery is fully depleted.  
The device continues to operate using the next available power source.    
External Battery Disconnected Info Message
 
This info message occurs when the external battery power source is 
lost and the device has switched to an alternate power source (either 
a detachable battery, if connected, or the internal battery if no other 
source is available). If external battery power returns, the ventilator 
will beep, but no message will appear on the display. 
Detachable Battery Disconnected Info Message
This info message occurs when the detachable battery power source 
is lost and the device has switched to an alternate power source (the 
internal battery if no other source is available). If detachable battery 
power returns, the ventilator will beep, but no message will appear 
on the display.
Start On Battery Info Message
This info message indicates that the ventilator has started on battery 
power and no AC power is available. The device operator should 
verify that this is what is wanted.   
Card Error Info Message 
 
This info message occurs when an unusable SD card is inserted into 
the ventilator. The device continues to operate but data cannot be 
logged onto the SD card.