Chapter 4 Ventilator Setup
59
Universal Porting Block
Exhalation Valve Line
Exhalation Valve Port
Proximal 
Pressure
Line
Proximal 
Pressure
Port
Exhalation Valve Port
on Universal Porting 
Block
Trilogy Universal Active 
PAP Tube Adapter
Refer to the instructions included with the Active Exhalation 
Device for more information.
c.  If your circuit type is Active Flow:
1. Connect the Flow Sensor to an active exhalation device as 
shown below.
Flow Sensor
2. Connect the other end of the flexible tubing to the active 
exhalation device. 
A.   Connect the active exhalation device to the flexible 
tubing that attaches to the ventilator.
B.  Connect the flow sensor to the patient interface (e.g., 
the tracheostomy tube).
Connecting an Active Device  
with Proximal Pressure
Connecting the Flow Sensor
Note: If the active exhalation 
device has the proximal pressure 
line connected when you are 
setting up an active flow circuit 
type, remove the proximal 
pressure line and cap the 
proximal pressure port before 
attaching to the flexible tubing.
Note: Make sure that 
components marked with an 
arrow are oriented properly.