C.  Connect the flow lines and the exhalation valve line to 
the Universal Porting Block on the ventilator as shown 
below.
 
Connect the White Striped Flow Line to 
the top, striped port on the Universal 
Porting Block on the device.
 
Connect the other Flow Line to the 
middle port on the Universal porting 
Block.
 
Connect the Exhalation Valve Line to the 
Exhalation Valve port on the top of the 
active exhalation device, and connect the 
other end of the line to the Exhalation 
Valve port on the Universal Porting Block.
Exhalation Valve Line
Flow Line
Exhalation Valve Port
on Porting Block
Universal Porting Block
on Trilogy Device
White Striped
Flow Line
Exhalation Valve Port
Refer to the instructions included with the active exhalation 
device for more information.
Connect a Water Trap 
If you are using an optional water trap, connect it to the patient 
circuit according to the manufacturer’s instructions.  
Note: When using an active 
exhalation device with proximal 
flow, ensure that any additional 
proximal ports are capped.
Note: The 
symbol 
appears next to the Exhalation 
Valve port on the Universal 
Porting Block to indicate where 
the active exhalation valve line 
connects.
Connecting the Active 
Exhalation Device with 
Proximal Flow Sensor