Triggering - The TRIGger
Subsystem
Subsystem
The TRIGger Subsystem allows you to configure how the instrument
reacts to incoming or outgoing triggers.
reacts to incoming or outgoing triggers.
Table 8 Triggering and Power Measurements
Hardware
Triggering
Triggering
Trigger Rearming
Software Triggering
Data Acquisition Functions
sens:func:stat
sens:func:stat
trig:inp
trig:inp:rearm
init:imm
init:cont
MINMax
LOGGing|STABility
IGNore
-
One power mea-
surement is per-
formed.
surement is per-
formed.
Automatically performs power
measurements.
measurements.
Automatically performs power
measurements until the func-
tion is finished.
measurements until the func-
tion is finished.
SMEasure
ON
Every hardware trigger starts a new power measure-
ment.
ment.
Every hardware trigger starts a
new power measurement until
the function is finished.
new power measurement until
the function is finished.
CMEasure
ON
The first hardware trigger starts
the function. Subsequent pow-
er measurements are automati-
cally performed until the
function is finished.
the function. Subsequent pow-
er measurements are automati-
cally performed until the
function is finished.
SMEasure
OFF
The first hardware trigger starts a new power mea-
surement. Further hardware triggers are ignored until
you send trig:inp:rearm again.
surement. Further hardware triggers are ignored until
you send trig:inp:rearm again.
Every hardware trigger starts a
new power measurement until
the function is finished.
new power measurement until
the function is finished.
CMEasure
OFF
The first hardware trigger starts
the function. Subsequent pow-
er measurements are automati-
cally performed until the
function is finished.
the function. Subsequent pow-
er measurements are automati-
cally performed until the
function is finished.