Installation 
22
309572ZAG
Fluid Temperature Sensor 
(FTS) Calibration 
(Non-RTD Versions)
1. Before turning on the unit, ensure all hoses 
and cables are properly connected. To 
ensure that the FTS in the hose is at the 
same temperature as the heaters, keep 
heat off and store the hose FTS near the 
machine for several minutes.
2. While holding down the temperature unit 
button (Fahrenheit - “F” or Celsius - “C”) 
turn the proportioner main power ON.
3. Hold the temperature unit button until tem-
perature is shown on the display. The fluid 
temperature sensor is now correctly cali-
brated.
Check Hoses for Leaks
1. Pressure check hose. See proportioner 
manual for priming instructions. 
2. After all lines are free of air, check for 
leaks. If there are leaks, relieve pressure 
as instructed in proportioner manual. 
3. Tighten connections, then pressurize again 
to ensure leaks have stopped. Relieve 
pressure.
Protective Covering
1. Wrap all fluid hose connections with 
electrical tape.
2. Fold FTS cable wire back on hose to 
ensure adequate strain relief. Wrap all 
electrical connections and cable connec-
tions with electrical tape to protect them 
from pulling apart and abrasion.
3. Install protective cover (see Accessories
page 31), or wrap hose bundle with duct 
tape to protect foam.
Calibrate the FTS ONLY at initial startup 
(the first time the unit is operated) and 
any time the hose length changes.
F
IG
. 14
Leave CAN cable connector (if used) 
accessible at whip hose joint for connec-
tor to the Remote Display Module Kit (if 
ordered).