Care and Maintenance
Table 10-1: Schedule for Periodic Maintenance
Removing Cooling Filter
Figure 10-4: Removing Cooling Filter
1. Remove black filter retaining bracket (1). (Pry off if necessary.)
2. Wash filter (2) in soapy water.
3. Dry thoroughly before reinstalling.
12,500 hours
• 12,500 hour preventive 
maintenance kit (P/N 
1001733). Kit contents are 
subject to change.
• Install 12,500 hour preventive 
maintenance kit
• Clean ventilator interior and exterior
• Complete performance verification 
procedure
CAUTION: The 12,500 hour preventive maintenance procedure is to be 
performed only by a qualified service technician.
As required
• External oxygen sensor
• Backup Battery
• Replace and recalibrate new sensor by 
running Extended Self-Test 
• For charging and maintenance 
CAUTION: