Respironics V200 Ventilator Operator’s Manual
8-9
Alarm 
Silence
Function: Disables the audio alarm for two minutes. When Alarm Silence is 
pressed before the end of a two-minute period, the two-minute timer is 
reset. Alarms that cannot be silenced are listed in Table 9-1: “Alarm Alert 
Messages” on page 9-5.
Indicator: Illuminated yellow when the audible alarm has been disabled; is 
active and stays on for two minutes when the ALARM SILENCE button is 
pressed. If ALARM SILENCE is active, and a new alarm condition occurs, 
which involves exceeding an active alarm limit, the visual alarm functions 
will be active. (Refer to “Alarm Silence” on page 9-2.) ALARM RESET 
clears ALARM SILENCE. If a medium or high priority alarm exists after 
ALARM RESET clears ALARM SILENCE, the audible alarm will begin.
Alarm 
Reset
Function: Clears the visual indicator for auto reset alarms, certain active 
alarms (see “Alarm Reset” on page 9-3), and reset of apnea ventilation 
back to the active mode of ventilation (see  “Apnea Ventilation” on page 8-
27). Al
arm Reset also terminates ALARM SILENCE.
100% O
2
Function: Press once to deliver 100% O
2
 to the patient for two minutes. 
Subsequent button presses will reset the timer to two minutes.
Indicator: Illuminated green; is active only when the 100% O2 front panel 
button has been pressed and the ventilator is delivering 100% O
2
 to the 
patient; remains on for the duration of 100% O
2
 delivery (two minutes).