 105 − 
 
6 F 8 A 0 9 3 4  
9.2.2    50 % Flow Rate Calibration 
Using Mag-Prover’s internal calibration circuit, 50% flow rate calibration can be performed. For 
calibration procedure, see the calibration procedure for 0% flow rate. (For 50% flow rate calibration, 
select "CAL 50" from the setting menu.) 
 
9.2.3    100 % Flow Rate Calibration (Span Calibration) 
Using Mag-Prover’s internal calibration circuit, 100% flow rate calibration can be performed. For 
calibration procedure, see the calibration procedure for 0% flow rate. (For 100% flow rate calibration, 
select "CAL 100%" from the setting menu.) 
 
9.2.4    Checking the Excitation Current 
 
You can monitor the exciting current value.   
 
●Checking the exciting current value 
Switch operation 
Display example 
Description 
 
R:CAL 
1 CAL   0% 
2 CAL  50% 
3 CAL 100% 
4 EXC DSPL
 
5 EXIT 
 
 
Select "EX DSPL" in the setting item 
selection screen. 
 
STEP1 
 
 
 
EXC DSPL 
 
   0.1998 
 
        A 
 
 
 
The excitation current value appears. 
 
Pushing       returns you to the setting 
menu. 
 
 
ESC
ESC 
ESC