 23 − 
 
6 F 8 A 0 9 3 4  
(3) Flow Direction 
 
Install
 
the
 
detector
 
in
 
accordance
 
with
 
the
 
flow
 
direction
 
arrow
 
on
 
the
 
detector.
 
See
 
Figure 
4.7.    If the actual flow runs opposite to the specified flow direction, the following 
display and output appears. 
 
For single range measurement, 
    ・LCD display:  
Instantaneous flow rate ------------- indicates negative values, 
 
 
 
Totalized flow ------------------------ no counts added. 
    ・Outut: 
 
Current output ------------------------ 4.0mA output, 
 
 
 
Pulse output --------------------------- No pulses 
For bidirectional range measurement, the flow in opposite direction results in a positive 
output value. See 10.3, “Multi-range Functions.” 
 
 
 
Figure 4.7 Flow direction arrow on the detector 
 
(4) Preventing an Empty Pipe Condition 
 
Design an upright pipe run (Figure 4.8) or sufficient head pressure (Fig. 4.9) at the 
downstream detector outlet if there is a possibility of the detector pipe becoming emptied. 
 
 
 
 
 
 
 
 
 
 
 
Figure 4.8 Detector with an upright pipe run at downstream outlet 
 
 
 
 
 
 
 
 
 
 
Figure 4.9 Detector with sufficient head pressure at downstream outlet 
Flow direction arrow 
Upright pipe run 
Detector 
Detector