8.2.12 Low Cutoff 
 
The low cutoff is the function to set the current output to zero forcefully if the flow rate is equal to or 
less than the low cutoff value set near 0%. 
 
The low cutoff value can be set within the range 0 to 10% in increments of 0.1%. 
 
You can check or change the low cutoff value as described below. 
 
The following is an example to change the low cutoff value from 1.0% to 3.0%. 
Switch operation 
Display example 
Description 
 
E:LOW CUT 
1 CUT VAL
 
2 DSPL SET 
3 EXIT 
 
 
Select "CUT VAL" from the setting item 
selection menu. 
 
STEP1 
 
 
 
CUT VAL 
 
      01.0 
 
         % 
 
 
 
The currently set low cutoff value (01.0% 
in this example) appears. 
 
Then push              . 
 
STEP2 
 
 
 
CUT VAL 
 
      
0
1.0 
 
         % 
 
 
 
The switches at the bottom change. 
(                                      are shown.) 
 
At the same time, the cursor appears.
 
 
STEP3 
 
 
 
CUT VAL 
 
      0
3
.0 
 
         % 
 
 
 
Push      to move the cursor to the 
desired digit and push      to change the 
number of the digit.
 
 
STEP4(=END) 
 
 
 
CUT VAL 
 
      03.0 
 
         % 
SET OK?
 
 
 
When you push      , a message appears 
to confirm whether the setting is OK or not. 
If OK, push            . If you want to redo the 
setting, push      . 
 
Pushing      cancels the setting operation 
and exits the setting screen. 
Note: If you try to set a value exceeding 10.0%, HIGH OVER SPEC error appears and the 
value returns to the previous value. Set a value again. 