Switch operation 
Display example 
Description 
STEP4(=END) 
 
 
 
DSPL SET 
 
LOW CUT 
 
        
SET OK?
 
 
 
When you push       , a message appears 
to confirm whether the setting is OK or not. 
If OK, push       . If you want to redo the 
setting, push       . 
 
Pushing      cancels the setting operation 
and exits the setting screen. 
Note: The measured value sent from the converter through communications is the value 
processed with display low cutoff. 
 
8.2.14 Still Water Zero Adjustment 
Zero adjustment is performed with the fluid held still in the detector's measurement pipe. 
Switch operation 
Display example 
Description 
 
F:ZERO 
1 ZERO ADJ
 
2 MANUAL 
3 EXIT 
 
 
Select "ZERO ADJ" from the   
configuration item selection menu.
 
 
STEP1 
 
 
 
ZERO ADJ 
 
       1.2 
 
         % 
 
 
 
The current flow rate measurement value 
appears. 
 
Then push and hold       longer. 
 
* Pushing       returns you to the setting 
menu. 
STEP2 
 
 
 
Longer 
ADJ READY 
 
       1.2 
 
         % 
        
 
 
The title display changes to ADJ READY, 
and the converter is ready for zero 
adjustment. 
 
* Pushing      returns you to the previous 
screen. 
 
STEP3 
 
 
 
NOW ZERO 
ADJUSTING 
 
          
        
 
 
Pushing       starts zero adjustment. 
 
STEP4(=END) 
 
 
 
ZERO ADJ 
 
       0.0 
 
         % 
 
 
 
Zero adjustment ends in several seconds and 
the flow rate measured value appears. 
 
Pushing       returns you to the setting 
menu. 
 
Note 1:  To start still water zero adjustment, push and hold              longer. 
Note 2:  Still water zero adjustment is possible only when the flow rate value is within the range of 
±1.25 m/s.
 
Note 3:  If you want to cancel the adjustment when ADJ READY is displayed, push       . This 
returns you to the state showing the flow rate measurement value on the screen. 