MR850 Technical Manual   *   Revision J   * Issued November 2005   *   Ref.  185041340 (185041713 internal)
 
30 
6.4.2 
Diagnostic Menu for Software Version 6.00 
Display  
Description 
HC 
Humidity Compensation (HC) algorithm 
Invasive mode, compensation range is 0.0 to 3.0 °C ( CSP = 37.0 to 40 °C ) 
Non-Invasive mode, compensation range is 0.0 to 5.0 °C ( CSP = 31.0 to 36.0 °C ) 
By pressing the Mute and Mode buttons together for 1 second or pressing the Mute 
and Power buttons together for 1 second the user can respectively move up or 
down through the settings listed below.  The humidifier will confirm the change with 
a double-beep. 
‘5.0’ = +5.0 °C of chamber compensation (Non-Invasive mode only) 
‘4.0’ = +4.0 °C of chamber compensation (Non-Invasive mode only) 
‘3.0’ = +3.0 °C of chamber compensation 
‘2.0’ = +2.0 °C of chamber compensation 
‘1.0’ = +1.0 °C of chamber compensation 
‘0.0’ = +0.0 °C of chamber compensation (NO compensation) 
‘-A-‘ = Automatic humidity compensation mode: 
                 Invasive mode: 0 to 3 °C in 0.5 °C steps 
                 Non-invasive mode: 0 to 3 °C in 0.5 °C steps 
CSP 
Chamber set point, in 0.1 °C resolution eg: 37.0 
Invasive mode range 35.5 to 40 °C. 
Non-Invasive mode range 31.0 to 36.0 °C. 
Cdc 
Chamber Duty Cycle (%) 
HP 
Heater Plate Temperature (1 °C) 
Flo 
Gas Flow Rate (0.1 LPM) 
  “---“ = Unknown Flow (flow measurement not started) 
FLr 
Gas Flow Rate Range: 
  “---“ = Unknown Flow (flow measurement not started) 
  “no” = No flow 
  “Lo” = Low flow, ( < 3 LPM ) 
  “In”  = Intermediate flow, ( 2 to 17 LPM ) 
  “Hi”  = High flow, ( > 13 LPM ) 
  “  -“  = Ventilated flow detected 
ASP 
Airway temperature set point (0.1 °C) 
Adc 
Airway Duty Cycle (%) 
LAS 
Last Alarm State, the display will blank, and the humidifier’s last alarm will be shown 
on the indicators.  To clear LAS, press and hold the mute and mode buttons for 1 
second.  The humidifier will beep when LAS has been cleared. 
LFS 
Last Fault State, refer to section 6.3.  To clear LFS, press and hold the mute and 
mode buttons for 1 second.  The humidifier will beep when LFS has been cleared. 
SoF Software 
version 
End 
Press the mute button to cycle to the start of the menu, or the diagnostic menu will 
automatically exit after 6 seconds