MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal)
30
6.4.2
Diagnostic Menu for Software Version 6.00
Display
Description
HC
Humidity Compensation (HC) algorithm
Invasive mode, compensation range is 0.0 to 3.0 °C ( CSP = 37.0 to 40 °C )
Non-Invasive mode, compensation range is 0.0 to 5.0 °C ( CSP = 31.0 to 36.0 °C )
By pressing the Mute and Mode buttons together for 1 second or pressing the Mute
and Power buttons together for 1 second the user can respectively move up or
down through the settings listed below. The humidifier will confirm the change with
a double-beep.
‘5.0’ = +5.0 °C of chamber compensation (Non-Invasive mode only)
‘4.0’ = +4.0 °C of chamber compensation (Non-Invasive mode only)
‘3.0’ = +3.0 °C of chamber compensation
‘2.0’ = +2.0 °C of chamber compensation
‘1.0’ = +1.0 °C of chamber compensation
‘0.0’ = +0.0 °C of chamber compensation (NO compensation)
‘-A-‘ = Automatic humidity compensation mode:
By pressing the Mute and Mode buttons together for 1 second or pressing the Mute
and Power buttons together for 1 second the user can respectively move up or
down through the settings listed below. The humidifier will confirm the change with
a double-beep.
‘5.0’ = +5.0 °C of chamber compensation (Non-Invasive mode only)
‘4.0’ = +4.0 °C of chamber compensation (Non-Invasive mode only)
‘3.0’ = +3.0 °C of chamber compensation
‘2.0’ = +2.0 °C of chamber compensation
‘1.0’ = +1.0 °C of chamber compensation
‘0.0’ = +0.0 °C of chamber compensation (NO compensation)
‘-A-‘ = Automatic humidity compensation mode:
Invasive mode: 0 to 3 °C in 0.5 °C steps
Non-invasive mode: 0 to 3 °C in 0.5 °C steps
Non-invasive mode: 0 to 3 °C in 0.5 °C steps
CSP
Chamber set point, in 0.1 °C resolution eg: 37.0
Invasive mode range 35.5 to 40 °C.
Non-Invasive mode range 31.0 to 36.0 °C.
Invasive mode range 35.5 to 40 °C.
Non-Invasive mode range 31.0 to 36.0 °C.
Cdc
Chamber Duty Cycle (%)
HP
Heater Plate Temperature (1 °C)
Flo
Gas Flow Rate (0.1 LPM)
“---“ = Unknown Flow (flow measurement not started)
“---“ = Unknown Flow (flow measurement not started)
FLr
Gas Flow Rate Range:
“---“ = Unknown Flow (flow measurement not started)
“no” = No flow
“Lo” = Low flow, ( < 3 LPM )
“In” = Intermediate flow, ( 2 to 17 LPM )
“Hi” = High flow, ( > 13 LPM )
“ -“ = Ventilated flow detected
“---“ = Unknown Flow (flow measurement not started)
“no” = No flow
“Lo” = Low flow, ( < 3 LPM )
“In” = Intermediate flow, ( 2 to 17 LPM )
“Hi” = High flow, ( > 13 LPM )
“ -“ = Ventilated flow detected
ASP
Airway temperature set point (0.1 °C)
Adc
Airway Duty Cycle (%)
LAS
Last Alarm State, the display will blank, and the humidifier’s last alarm will be shown
on the indicators. To clear LAS, press and hold the mute and mode buttons for 1
second. The humidifier will beep when LAS has been cleared.
on the indicators. To clear LAS, press and hold the mute and mode buttons for 1
second. The humidifier will beep when LAS has been cleared.
LFS
Last Fault State, refer to section 6.3. To clear LFS, press and hold the mute and
mode buttons for 1 second. The humidifier will beep when LFS has been cleared.
mode buttons for 1 second. The humidifier will beep when LFS has been cleared.
SoF Software
version
End
Press the mute button to cycle to the start of the menu, or the diagnostic menu will
automatically exit after 6 seconds
automatically exit after 6 seconds