MR850 Technical Manual   *   Revision J   * Issued November 2005   *   Ref.  185041340 (185041713 internal)
 
32 
6.4.4 
Diagnostic Menu for Software Version 7.14 
Display 
Description 
HC 
Humidity Compensation (HC) algorithm 
Note: HC is inactive while operating under non-heater wire control. 
Invasive mode, compensation range is 0.0 to 5.0 °C ( CSP = 37.0 to 42 °C ) 
Non-Invasive mode, compensation range is 0.0 to 5.0 °C ( CSP = 31.0 to 36.0 °C ) 
By pressing the Mute and Mode buttons together for 1 second or pressing the Mute and Power buttons 
together for 1 second the user can respectively move up or down through the settings listed below.  The 
humidifier will confirm the change with a double-beep. 
‘5.0’ = +5.0 °C of chamber compensation 
‘4.0’ = +4.0 °C of chamber compensation 
‘3.0’ = +3.0 °C of chamber compensation 
‘2.0’ = +2.0 °C of chamber compensation 
‘1.0’ = +1.0 °C of chamber compensation 
‘0.0’ = +0.0 °C of chamber compensation (NO compensation) 
‘-A-‘ = Automatic humidity compensation mode: 
Invasive mode: 0 to 5 °C in 0.5 °C steps 
Non-invasive mode: 0 to 5 °C in 1 °C steps 
Cct 
Connected breathing circuit identification: 
  “S“ = Standard inspiratory heater connected 
  “C” = Coaxial inspiratory heater connected 
  “E” = Expiratory heater connected 
  “---” = No heaters detected while under heater wire control 
CSP 
Chamber set point, in 0.1 °C resolution eg: 37.0 
Invasive mode range 35.5 to 42 °C. 
Non-Invasive mode range 31.0 to 36.0 °C. 
Cdc 
Chamber Duty Cycle (%) 
During selection of this menu and while a chamber probe out test is active the chamber probe indicator will 
light. 
CHP 
Chamber Power / Flow ratio (W/LPM) 
hP 
Heater Plate Temperature (1 °C) 
Flo 
Gas Flow Rate (0.1 LPM) 
  “---“ = Unknown Flow (flow measurement not started) 
FLr 
Gas Flow Rate Range: 
  “---“ = Unknown Flow (flow measurement not started) 
  “no” = No flow, ( Stand-by ) 
  “Lo” = Low flow, ( < 3 LPM ) 
  “In”  = Intermediate flow, ( 2 to 17 LPM ) 
  “Hi”  = High flow, ( > 13 LPM ) 
  “  -“  = Ventilated flow detected 
ASP 
Airway temperature set point (0.1 °C) 
Adc 
Airway Duty Cycle (%) 
During selection of this menu and while an airway probe out test is active the airway probe indicator will light. 
H2O 
Water out number, used to detect the presence of chamber water, calculated from, chamber power / (heater 
plate temp. – chamber temp.). 
During selection of this menu and while the water out number falls below a dry chamber threshold the water 
out indicator will light. 
LAS 
Last Alarm State, the display will blank, and the humidifier’s last alarm will be shown on the indicators.  To 
clear LAS, press and hold the mute and mode buttons for 1 second.  The humidifier will beep when LAS has 
been cleared. 
LFS 
Last Fault State, refer to section 6.3.  To clear LFS, press and hold the mute and mode buttons for 1 second.  
The humidifier will beep when LFS has been cleared. 
SoF 
Software version 
End 
Press the mute button to cycle to the start of the menu, or the diagnostic menu will automatically exit after 6 