fisher--paykel mr850 Manual De Usuario

Descargar
Página de 62
 
MR850 Technical Manual   *   Revision J   * Issued November 2005   *   Ref.  185041340 (185041713 internal)
 
Expiratory Limb  
The section of the breathing circuit that takes the expired gases from 
the patient. 
1.3 Definitions 
NOTE: A NOTE provides important information or explanation of procedures or conditions which 
may otherwise be misinterpreted or overlooked. 
CAUTION: A CAUTION statement designates the possibility of damage to this or other 
equipment if a procedure is not followed exactly. 
WARNING: 
A WARNING statement refers to conditions with a possibility of personal injury if a 
procedure is not followed exactly.
 
1.4 Product 
Application 
The MR850 is a respiratory humidifier designed for use in hospital intensive care units.  It is 
used to provide optimum humidity to respiratory gases delivered to patients via endotracheal  
tubes or face masks. 
Refer to the Respiratory Humidification Product Catalogue or your local Fisher & Paykel 
Healthcare representative for a list of approved accessories. 
WARNING: 
• 
The use of breathing circuits, chambers or other accessories which are not approved 
by Fisher & Paykel Healthcare may impair performance or compromise safety. 
• 
Ensure that Invasive mode is set for patients that have bypassed airways. 
• 
Ensure maintenance of grounding integrity by connection to a "hospital grade" 
receptacle. 
• 
Always disconnect supply before servicing. 
• 
When mounting a humidifier adjacent to a patient ensure that the humidifier is always 
securely mounted and positioned lower than the patient. 
• 
The operation of high frequency surgical apparatus, shortwave or microwave 
equipment in the vicinity of the humidifier may adversely affect its function.  If this 
occurs, the humidifier should be removed from the vicinity of such devices. 
• 
Ensure that both temperature probe sensors are correctly and securely fitted.  Failure 
to do so may result in gas temperatures in excess of 41 °C being delivered to the 
patient. 
• 
Do not touch the glass tip of the chamber temperature probe during use.  Keep black 
connectors dry at all times. 
• 
Visually inspect accessories for damage before use. 
• 
Normal operation can not be guaranteed if powered from a source other than a pure 
sine wave, such as a square wave inverter.