MR850 Technical Manual * Revision J * Issued November 2005 * Ref. 185041340 (185041713 internal)
10
3 Specifications
3.1 Mechanical
Dimensions:
140 mm x 173 x 135 (without chamber fitted)
Weight:
2.8 kg (without chamber fitted)
Approx. 3.1 kg (with chamber fitted, and filled with water)
Approx. 3.1 kg (with chamber fitted, and filled with water)
3.2 Electrical
MR850 Model Number
Supply Voltage
Supply Current
MR850Axx
MR850Pxx
MR850Jxx
MR850Gxx
MR850Pxx
MR850Jxx
MR850Gxx
230 V~
127 V~
115 V~
100 V~
127 V~
115 V~
100 V~
1.0 A Max
1.8 A Max
2.0 A Max
2.4 A Max
1.8 A Max
2.0 A Max
2.4 A Max
Supply Frequency:
50 or 60 Hz, Sine Wave
Heater Plate Capacity
150 W at nominal mains voltage
HP Thermal Cutout:
118 ± 6 °C
Heater Wire Supply:
22 ± 5 V~, 2.73 A Max, 50 or 60 Hz
Maximum Heater Wire Load:
8.0 Ω.
3.3 Temperature
Range
3.3.1 Heater
Wire
Mode
Invasive Mode:
Chamber Set Point: 35.5 to 37 °C
Chamber Set Point: 35.5 to 40 °C (versions 5.33, 5.34, 5.45, 5.70,
6.00, 7.00, 7.21).
Chamber Set Point: 35.5 to 42 °C (versions 7.14, 7.17 & 7.22).
Airway Set Point: 35 to 40 °C
Chamber Set Point: 35.5 to 40 °C (versions 5.33, 5.34, 5.45, 5.70,
6.00, 7.00, 7.21).
Chamber Set Point: 35.5 to 42 °C (versions 7.14, 7.17 & 7.22).
Airway Set Point: 35 to 40 °C
Non-Invasive Mode:
Chamber Set Point: 31 °C
Chamber Set Point: 31 to 34 °C (versions 5.33, 5.34, 5.45, 5.70,
6.00).
Chamber Set Point: 31 to 36 °C (versions 7.00, 7.14, 7.17, 7.21,
7.22).
Airway Set Point: 28 to 34 °C
Chamber Set Point: 31 to 34 °C (versions 5.33, 5.34, 5.45, 5.70,
6.00).
Chamber Set Point: 31 to 36 °C (versions 7.00, 7.14, 7.17, 7.21,
7.22).
Airway Set Point: 28 to 34 °C
3.3.2
Non Heater Wire Mode
Invasive Mode:
Airway Set point: 37 °C (chamber temperature limited to 66 °C)
Non-invasive Mode:
Airway Set point: 31 °C (chamber temperature limited to 66 °C)
Display:
Three digit, 14 mm, 7 segment LED
Range: 10 to 70 °C
Range: 10 to 70 °C
Accuracy:
± 0.3 °C (in 25 to 45 °C temperature range)
3.3.3 Alarm
Parameters
High Temperature Alarm:
Causes an immediate, audible and visible alarm at a displayed
temperature of 41 °C or if the airway temperature exceeds 43 °C
(see section 4.3)
temperature of 41 °C or if the airway temperature exceeds 43 °C
(see section 4.3)
Temperature Alarm:
Invasive Mode:
After 10 minutes @ 29.5 °C causes an audible and visible alarm.
After 60 minutes @ 34.5 °C causes an audible and visible alarm
(see section 4.5)
NOTE: The temperature indicator lights if the displayed
temperature drops below 35.4 °C, initially providing a temperature
warning.
After 10 minutes @ 29.5 °C causes an audible and visible alarm.
After 60 minutes @ 34.5 °C causes an audible and visible alarm
(see section 4.5)
NOTE: The temperature indicator lights if the displayed
temperature drops below 35.4 °C, initially providing a temperature
warning.