GE PT878GC Manual De Usuario

Descargar
Página de 308
3-11
April 2004
Installing the Transducers
1. Apply a bead of couplant 6 mm (0.25 in.) wide along the 
entire length of each transducer face, as shown in Figure 3-10 
below. 
Figure 3-10:  Couplant on Transducer Face
Note: Do not slide the transducer with couplant along the 
surface of the pipe when mounting the transducer.
2. Set the first mounting block (either left edge or right edge) at a 
convenient number on the scale, such as 1 in. or 1 cm. Install 
the first transducer with the BNC connector pointing away 
from the center of the V block fixture. Tighten the transducer 
mounting thumbscrew onto the slider, which in turn applies 
pressure to the transducer. Use a handtight grip to set the 
transducer in contact with the pipe, as shown in Figure 3-11 
below. Use a wrench to tighten the backing nut to prevent 
loosening due to vibration and thermal expansion. 
IMPORTANT:  Do not use a wrench or pliers on the 
thumbscrew.
Figure 3-11:  Installing the First Transducer