Installing the V Series Clamping Fixture 
and Transducers
Note: A complete installation involves the clamping fixture, 
transducers and dampening material. If you are using 
dampening material, refer to 
Installing Dampening 
Material on page 3-21.
To install the V Series clamping fixture and transducers, 
complete the following steps:
Installing the Fixture
1. Position the half of the clamping fixture with the threaded 
rods around the pipe, as shown in Figure 3-7 below. Orient the 
fixture in the 3 o’clock position on a horizontal pipe. 
2. Position the mating half of the fixture over the threaded rods 
in the 9 o’clock position. Figure 3-7 below shows the two 
mounted halves. 
Note: If you are using a V4 fixture with a 1.5 to 2-in. pipe, GE 
Panametrics suggests removing the nut and the mounting 
bracket, rotating it 180° and replacing the bracket and 
nut for a more secure fit.
Figure 3-7:  Mounting the Two Halves of the Fixture