Amplifier Characteristics
Open thermocouple detection current.....250 nA max
1
(assuming floating thermocouple)
Common-mode rejection ratio................110 dB min
Output range
...........................................±10 V max
Transfer Characteristics
Gain ........................................................100
Gain error
................................................±0.08% max
Gain stability
...........................................±0.0005%/°C max
2
Offset error
.............................................±5 µV max (post calibration)
Offset stability
........................................±0.6 µV/°C max
Nonlinearity
............................................±0.004% max
Recommended warm-up time.................5 minutes
Measurement Accuracy 
3
1   
Temperature range is 23 °C 
±5 °C.
2   
Temperature range is 0 to 50 °C.
3   
Total system measurement error for operating temperature within 
±5 °C of calibration temperature. Includes 
PCI/AT-MIO-16XE-50 one-year accuracy specification of 0.01% 
±412 µV. SCC-TC01/02 accuracy specification of 0.08% 
±5 µV, and reference junction measurement accuracy of 0.5 °C. Assumes averaging. Non-averaged, single-point reading has 
an additional uncertainty (up to 
±0.1 °C for J-type thermocouple).
Thermocouple 
Type
Temperature 
Range (
°C)
Maximum 
Inaccuracy (
±°C)
Typical 
Inaccuracy (
±°C)
B
400 to 600
600 to 1,800
2.5
2
0.31
0.13
E
–200 to –100
–100 to 600
600 to 1,000
2
1
1.5
1.76
1
0.6
J
–200 to –100
–100 to 500
500 to 1,100
2
1
1.5
1.6
0.9
0.72