National Instruments SCC-TC01 Manual De Usuario

Descargar
Página de 20
SCC-TC Series Thermocouple Input Modules User Guide
6
ni.com
The amplified thermocouple signal and cold-junction sensor signal 
are measured by E Series DAQ device channel X and channel X
+8 
respectively, where X is 0 to 7 depending on where you plug the 
SCC-TC0X. Refer to Figure 3 for SCC-TC0X signal connections.
Figure 3.  SCC-TC0X Signal Connections
Cold-Junction Sensor Output and Accuracy
The cold-junction sensor voltage output varies from 1.91 to 0.58 V over a 
0 to 55 
°C temperature range.
Note
NI-DAQ includes thermistor conversion utilities that implement the equations listed 
below. Refer to your software documentation for more information on these utilities.
Thermocouple
Source
ACH(
)
ACH(
X
+8)
SCC-TC0
X
E Series 
DAQ Device
+
4.75 k
Ω 
1%
5 k
Ω 
0.1%
10 k
Ω 
10 k
Ω 
2.5 k
Ω 
+
5 k
Ω 
at 25
o
 
C
10 M
Ω 
10 M
Ω 
10 
µF
16 V
0.1 
µF
0.1 
µF
-t
o
Ref 5 V
Ref 5 V
LM 4040
2.5 V
0.1%
2-Pole Filter/
Buffer Stage
Offset
Calibrator
AISENSE
TC+
GND
(SCC-TC02
 only)
TC-
AIGND
+