WARNING 
Turn off the Doppler before removing or replacing the battery. 
Replace  alkaline  batteries  with  those  of  identical  specifications 
provided 
by 
the 
manufacturer 
or 
purchased 
locally.SeeChapterProduct  Specificationsfor  details  about  battery 
specifications. 
If  the  batteries  have  been  inserted  incorrectly,  the  Doppler  will  not 
function or it will be damaged. 
Do not disassemble or short-circuit batteries. 
Do not recharge batteries. 
Do not dispose of batteries in fire or water. 
Do not allow metal objects to contact the battery terminals. 
Do  not  mix  with  used  or  other  battery  type  (such  as  alkaline  with 
carbon zinc). 
Do  not  solder  the  batteries  directly.  If  soldering  or  welding 
connection to the battery is required, consult our engineer for proper 
methods. 
10 
Do not over-discharge batteries. 
11 
To install or remove batteries, follow the equipment manufacturer‘s 
instructions. 
12 
Keep  battery  away  from  small  children.  If  swallowed,  consult  a 
physician at once. 
13 
Store the battery in cool, dry place before use.Do not keep batteries 
at temperature of 45°C or above, or at humidity of 75% or above. 
14 
Dispose  the  battery  according  to  the  local  regulations.  Refer  to 
IEC61429 for standard disposal when necessary. 
Switching On 
Touch  the  On/Off  touch  key  for  about  1second  when  the  Doppler  is  off, 
and  the  Doppler  will  display  the  switching  on  interface
before 
switching to display the test interface
Switching Off 
Touch  the  On/Off  touch  key  for  about  1second  when  the  Doppler  is  on, 
and the Doppler will be switched off. 
If the Doppler is not in operation or no signal is received for 2 minutes, the 
Doppler will switch off automatically. 
  FHR Monitoring 
Before  applying  the  Doppler  to  inspect  FHR,  you  should  always  check 
whether  the  Doppler  is  in  good  condition  and  whether  there  is  evident 
damage  that  might  affect  patient‘s  safety  and  the  device‘s  function.If 
evident  damage  is  found,  stop  using  it  at  once  and  replace  it  with  a  good 
one. 
Procedures to Monitor FHR: 
a) 
Have the patient lie on her back. 
b) 
Apply 
appropriate 
amount of coupling gel 
to 
the 
ultrasonic 
transducer  head  of  the 
Doppler  and  switch  on 
the Doppler. 
c) 
Palpate  the  patient‘s 
abdomen 
gently 
to 
confirm  the  fetus‘s 
position.   
d) 
Place  the  Doppler  on 
the  patient‘s  abdomen, 
and  move it around the 
fetus‘s position or tilt it 
until 
clear 
and 
rhythmic heart sound is 
heard and FHR numeric is stably displayed. 
Note:
 
1  Do  not  mistake  the  maternal  heart  rate  for  fetal  heart  rate.Do  not 
mistake  the  maternal  heart  rate  for  fetal  heart  rate.  The  fetal  pulse 
should be different from the maternal pulse, which can be measured at 
the wrist or neck.
 
2  Do not wear gloves to touch the keys. If there's water and coupling gel 
on  the  fingers,  please  clean  them  first  or  the  touching  effect  will  be 
influenced. 
How to Find the Best FH Signal: 
1) 
The easiest way: take the position the doctor last monitored for FHR 
as  a  reference  and  move  the  Doppler  around  the  position  slowly 
until the best FH signal is found. 
2) 
The  fetal  heart  position  may  change  as  the  fetus  moves  inside  the 
uterus.  You  can  confirm  the  fetal  position  first  according  to  the 
position of the uterus fundus in different gestational weeks. 
The  clearest  and  loudest  fetal  heart  sound  is  generally  obtained 
when  the  Doppler  is  placed  on  the  fetus‘s  back.  Fetal  movement  is 
usually the movement of fetal limbs. So, if frequent fetal movement 
occurs  at  the  right  side  of  the  abdomen,  the  fetus‘s  back  is 
probablyat  the  left  sideand  vice  versa.You  can  find  the  fetus‘s  back 
according to fetal movement‘s position. 
If the fetus is in cephalic delivery position, the fetal heart is either on 
the  right  side  or  on  the  left  side  below  the  navel;  if  the  fetus  is  in 
breech delivery position, the fetal heart  is either on the right side or 
on the left side above the navel. 
Steps to Find Fetal Heart: 
Have the patient lie on back and relax >> confirm fetal position by hand >> 
apply  coupling  gel  to  the  Doppler>>  place  the  Doppler  on  patient‘s 
abdomen  and  start  looking  for  the  fetal  heart  >>  the  fetal  heart  is  found 
when 
the 
Doppler 
gives 
out 
continuing 
thumping 
sound 
―boom-boom-boom‖. 
 
CAUTION 
The Doppler‘s degree of protection against harmful ingress of water 
is IP22. Do not immerse it in water. 
The Doppler is delicate and sensitive. Please handle it with care and 
try  to  avoid  dropping  on  to  the  ground  or  any  hard  surfaces.  Any 
damage caused by dropping is not covered by the warranty. 
Keep  the  coupling  gel  away  from  children.  If  swallowed,  consult  a 
physician at once. 
Note: 
1  The best quality of fetal heart signal is obtained only when the Doppler 
is placed on the best monitoring position. 
2  Do  not  place  the  Doppler  near  positions  where  placental  sound  or 
umbilical blood flow sound is loud. 
3  If  the  fetus  is  in  the  cephalic  position  and  the  mother  is  supine,  the 
clearest  heart  sound  will  normally  be  found  on  the  midline  below  the 
navel.  During  monitoring,  the  pregnant  woman‘s  prolonged  lying  in 
the  supine  position  should  be  avoided  to  reduce  the  possibility  of 
supine  hypotension.  Putting  a  pillow  or  cushion  under  the  patient‘s 
head or feet can be of help. 
4  It  is  not  possible  to  obtain  accurate  FHR  unless  a  clear  fetal  heart 
signal  is detected. If the  calculated FHR is not in accordance  with the 
beat  of  the  fetal  heart  sound,  the  fetal  heart  sound  auscultation  result 
shall prevail. 
5  When applied to the patient, the Doppler may warm slightly (less than 
2°C  (35.6°F)  above  ambient  temperature).  When  NOT  applied,  the 
Doppler  may  slightly  (less  than  5°C  (41°F)  above  ambient 
temperature). 
After Monitoring 
1)  Switch off the Doppler. 
2)  Wipe the remaining  gel off the patient and the probe  with a clean soft 
cloth or tissue. 
  Mobile Application Software (APP)   
SD1  can  connect  to  mobile  phones  with  its  Bluetooth  function  (optional). 
The SD1 APP has both Android and iOS versions. 
iOS APP operating environment: 
Android APP operating 
environment: 
A) hardware environment 
A) hardware environment 
Processor: dual-core Apple A6 
CPU: frequency≥1.0GHz 
RAM: ≥1GB 
RAM: ≥1GB 
B) software environment: iOS 
8.0 and above operating system 
B)software environment: Android 
4.3 and above operating system 
C)network environment: support 
Bluetooth 
C)network environment: support 
Bluetooth 
How to use SD1 Medical APP 
1.Download and install software 
Scan either of the following QR codes to download the SD1Medical APP, 
and install and run it as prompted. 
TBC 
TBC 
Note: 
1  Your  mobile  phone  may  prohibit  the 
installation  of  ―applications  from 
unknown sources‖. Enter Settings to 
allow the installation first. 
2  For  normal  functioning  of  the  APP, 
please  give  the  APP  function-related 
permissions. 
3  For  how  to  use  the  APP,  read  the 
instructions 
in 
the 
About 
sub-interface  under  the  Settings 
interface of the APP. 
2.Activate the device 
Open the APP and go to Settings>Activation and input SD1 activation 
code (14 numbers after 01). 
3.Pair device 
Open Bluetooth function of the mobile to automatically pair the SD1. 
4.Start detection 
Put the coupling gel on SD1 and position the probe to the optimal place of 
maternity's abdomen. And click the "start" key.
 After pressing start, 
confirm that the data on the APP and the SD1 probe match. As with 
any Bluetooth communication, it is important to make sure the 
connection is not compromised.
 
5.Adjust the fetal heart beat sound volume 
When using mobile phone to play the fetal heart beat sound, you can adjust 
the volume with the volume keys of the mobile phone. When using SD1 to 
play the heart beat sound, touch ‗volume+‘ or ‗volume -‘ to adjust the 
volume. 
6.Finish the monitoring 
When the monitoring is finished, click ‗Stop‘ touch key and the detection 
data will be saved automatically. 
Note:Please make sure your mobile phone has enough battery power,and 
avoid killing the process directly or switching to other applications during 
the fetal heart monitoring. 
7.Real time detection mode and DEMO mode 
We provide DEMO mode for users' reference. You can turn on DEMO key 
in Setup and enter fetal heart monitoring interface to watch the DEMO. 
The word ‗DEMO‘ is displayed in the interface to distinguish from real 
time detection. 
WARNING 
SD1  complies  with  Part  15  of  the  FCC  Rules.  Operation  is  subject  to  the 
following two conditions:   
1)this device may not cause harmful interference, and   
2)this device must accept any interference received, including interference 
that may cause undesired operation. 
NOTE: 
1.This equipment (SD1) has been tested and found to comply with the 
limits for a Class B digital device, pursuant to part 15 of the FCC Rules. 
These limits are designed to provide reasonable protection against harmful 
interference in a residential installation. This equipment generates, uses 
and can radiate radio frequency energy and, if not installed and used in 
accordance with the instructions, may cause harmful interference to radio 
communications. However, there is no guarantee that interference will not 
occur in a particular installation. If this equipment does cause harmful 
interference to radio or television reception, which can be determined by 
turning the equipment off and on, the user is encouraged to try to 
correctthe interference by one or more of the following measures: 
-Reorient or relocate the receiving antenna. 
-Increase the separation between the equipment and receiver. 
-Connect the equipment into an outlet on a circuit different from that to 
which the receiver is connected. 
-Consult the dealer or an experienced radio/TV technician for help. 
2.Any changes or modifications to this unit not expressly approved by the 
party responsible for compliance could void the user's authority to operate 
the equipment. 
  Maintenance and Cleaning 
Maintenance   
Before  each  use,  check  if  the  equipment  has  visible  evidence  of  damage 
that  may  affect  the  patient  and  the  operator‘s  safety  or  the  Doppler‘s 
functioning. If the damage is evident, contact the manufacturer for service 
or replace it. 
The  overall  check  of  the  Doppler,  including  safety  check  and  function 
check,  should  be  performed  by  qualified  personnel  every  12  months,  and 
each time after service. And safety check must include current leakage test 
and  insulation  test.  Besides  the  above  requirements,  comply  with  local 
regulations on maintenance and measurement. 
The accuracy of FHR is determined by the Doppler and cannot be adjusted 
by user. If you have doubt concerning the accuracy of FHR,  verify it with 
other  methods  such  as  using  a  stethoscope,  or  contact  local  distributor  or 
the manufacturer for help. 
The  Doppler  is  frangible  and  must  be  handled  with  care.Wipe  the 
remaining  gel  off  the  Dopplerafter  each  use.  These  measures  can  help 
prolong the Doppler‘s life. 
Replace the accessories such as the battery according  to use. If any of the 
accessories  are  damaged,  refer  to  chapter  Ordering  Information  for 
details and order new ones. 
Please  check  the  label  for  the  date  of  manufacture,  the  service  life  is  5 
years  (The  service  life  is  limited  to  the  Doppler,  not  including  the 
replaceable accessories. The  only replaceable accessory of SD1 is battery. 
The frequency of usage is 8 hours/day). 
Cleaning 
Before cleaning, switch off the Doppler. 
Keep the exterior surface of the device clean and free of dust and dirt.   
Clean the exterior surface of the Doppler with a dry, soft cloth. If necessary, 
clean  it  using  a  soft  cloth  dampened  with  mild  near  neutral  detergent, 
ethanol (75%) or isopropanol (70%), and then wipe it dry with a dry cloth 
immediately. 
CAUTION 
1  Do not use strong solvent, such as acetone. 
2  Never use an abrasive such as steel wool or metal polish. 
The Doppler‘s degree of protection against harmful ingress of water is 
IP22. Do not immerse it in water. 
4  Do not remain any solution on the surface after cleaning. 
Disinfection 
In  normal  use  theDoppler  does  not  need  disinfection.  In  case  of  being 
soiled,  clean  the  main  unit  case  and  then  disinfect  it  by  wiping  it  with  a 
soft cloth dampened with ethanol (75%) or isopropanol (70%). Then wipe 
it dry with a dry cloth. 
CAUTION 
Do not immerse the Doppler into the disinfector. 
Sterilization 
Do not sterilize the Doppler. 
NOTE: 
After  cleaning  or  disinfection,  check  if  the  Doppler  functions  well.  If  any 
problem  is  detected,  please  contact  the  manufacturer  for  service  before 
reusing it. 
Checking Item 
Method 
Visual Check   
Inspect the Doppler for any damage. 
Functional 
Check 
Check  if  the  Doppler  can  be  switched  on  and  off 
normally  (see  Switching  On  and  Switching 
Off)
When the Doppler is switched on, check if the 
display  panel  works  as  described  in  LCD 
Display&Touch  Keys
;  touch  the  ultrasonic 
transducer head gently with your hand and check if 
the Doppler gives out sound normally. 
  Product Specifications 
Product Information 
Product Name 
Ultrasonic Pocket Doppler 
Model 
SD1 
Complied Standards 
IEC 
60601-1:2005/A1:2012, 
EN 
60601-1:2006/A1:2013, 
IEC 
60601-1-2:2014, IEC 60601-2-37:2015,IEC 61266:1994 
Classification 
Anti-electric Shock Type: 
Internally powered equipment 
Anti-electric Shock Degree: 
Type BF equipment 
Degree 
of 
Protection 
against 
Harmful Ingress of Water: 
IP22. Do not immerse it in water. 
Degree  of  Safety  in  Presenceof 
Equipment  not  suitable  for  use  in 
The 
coupling 
gel 
should  not  exceed 
this limit.
 
This  area  can 
be immerged in 
coupling gel