BAROSTIM NEO LEGACY REFERENCE GUIDE 
8-4
 
Parameter 
Description 
Range of Values 
Burst Enable 
Determines whether therapy pulses are applied throughout 
the burst cycle in a continuous manner or if a cycle of 
active and rest periods is being used. 
On, Off 
Burst Duration 
Determines the length of the active portion of the burst 
cycle during which the Therapy Frequency is delivered. 
NOTE: This parameter is not shown if Burst Enable is Off. 
50 milliseconds to 1950 
milliseconds 
Burst Interval 
Determines the total length of the burst cycle including the 
active portion and the rest portion. 
NOTE: This parameter is not shown if Burst Enable is Off. 
100 milliseconds to 2000 
milliseconds 
The Compliance column of each therapy provides an indication of the Compliance of each pathway.  
Compliance is measured when therapies are tested on the “Edit and Test…” screen.  The  Compliance 
indication can be one of the following: 
“---” 
No Compliance measurement has been taken 
” 
IPG is able to deliver programmed amplitude 
” 
IPG is not
Compliance is a measure of the ability of the IPG to deliver the programmed output amplitude.  The IPG 
delivers output pulses that maintain a constant electrical current during the duration of each pulse.  There 
are situations in which the device cannot maintain this constant current.  When this occurs, the device will 
not deliver the programmed amplitude throughout the duration of the pulse  and Compliance will fail.  
Settings that are out of Compliance cannot be saved in the IPG. 
 able to deliver programmed amplitude 
NOTE: All therapies with pathways enabled must be evaluated for Compliance at each follow-up 
so that the proper settings are sent to the IPG.  This should be performed even if the therapy 
settings are not being changed. 
Edit and Test Screen 
The Edit and Test Screen is used to adjust therapy settings and record therapy effectiveness.  Follow these 
steps to adjust, assess, and program therapy settings: 
1.  Select desired therapy settings and test mode in the “Therapy N” window. 
2.  Click the Test Now button to start delivery of the desired settings. 
3.  Use the Elapsed Time to determine when to perform a blood pressure measurement. 
4.  If testing is complete, click “Record” to add the entry to the Patient Response Log. 
5.  If more testing needs to be performed, go back to Step 1. 
6.  Enter BP, Heart Rate, and any observational notes on any entry in the list can be entered at any 
time by double clicking on the box containing the information. 
7.  Select the Patient Response Log entry containing the final desired settings for the therapy. 
8.  Click “Save Selected Log Entry as Therapy N” to program the settings for Therapy N. 
The “Therapy N” window contains a group of parameter settings  that can be used to evaluate therapy 
efficacy.  Use the Pathway check boxes to control which channels will be assessed.  Adjust Pulse Width, 
Amplitude, Frequency, and Burst parameters to the desired settings. 
Click Test Now to transmit the settings in the Therapy N window to the IPG.  The settings are transferred 
to the Pulse Parameters section of the Patient Response Log window and the Elapsed Time is restarted.  If a