BAROSTIM NEO LEGACY REFERENCE GUIDE 
8-5
 
test is currently in progress, these current settings and compliance results will be added to the Patient 
Response Log when Test Now is pressed. 
Click Stop Test to stop therapy, allowing observation of the patient response without therapy.  Therapy is 
disabled as reflected in the Pulse Parameters section of the Patient Response Log window.    A Patient 
Response Log entry is automatically added with a note indicating the therapy was stopped.  The Elapsed 
Time is restarted. 
The Patient Response Log window allows monitoring of Compliance during  the test and provides the 
Elapsed Time since the settings were invoked.   
NOTE:  If a Compliance check fails when using aggressive therapy settings, first adjust the 
Amplitude downward until the Compliance check passes.  If further reduction of Amplitude is not 
desired, reducing Pulse Width may allow the Compliance check to pass.  Also, be sure the lead 
impedance for the channel being tested is in a normal range. 
The Patient Response Log window also provides an estimation of the battery life utilizing the parameters 
being tested.  This estimate does not take the Therapy Schedule into account.  Instead, therapies under test 
always assume a 24 hour schedule.  This allows comparison of the current settings to other settings that 
have been tested.  However, these estimates should not be compared to the Battery Life estimate in the IPG 
Status section at the top of the screen.  The Battery Life estimate in IPG Status always represents the 
longevity of the currently programmed permanent parameter settings. 
Clicking the Record button adds an entry to the Patient Response Log.  An entry can be selected by clicking 
somewhere on the desired  entry in the log.  The selected entry is highlighted.  A vertical scroll bar will 
appear to the right of the entry list if there are more entries than can be displayed.  Use the scroll bar to find 
the desired entry if it isn’t visible.  Measured results for BP and Heart Rate, as well as any observational 
notes that are related to the therapy settings, can be entered by clicking on the box of interest and entering 
the information.  (Note that the on-screen keyboard can be used to aid data entry.)  The selected entry can 
be removed by clicking the 
  button.  All edits and deletions are permanent.  Click the Use “Selected Log 
Entry for Test” button to load the settings of the selected Patient Response Log entry into the Therapy N 
window.  This provides convenient access to prior settings if more testing is desired. 
There are two options to return to the Main Screen. 
•  Select the permanent therapy  settings by choosing a Patient Log Entry that contains the desired 
settings and also indicates that Compliance has passed on all active pathways.  Then press the 
“Save Selected Log Entry as Therapy N” button. 
•  If no permanent changes are to be made, choose “Exit without Saving”. 
NOTE:  All therapies used in  the  Therapy Schedule should be evaluated for compliance and re-
saved as the permanent therapy if the compliance result has not passed. 
Adjusting the Schedule 
The Schedule controls the portion of each day during which therapy is delivered.  The Schedule is a 24 hour 
clock that is started at midnight each day.  The current Schedule settings are displayed in the Schedule 
window (see Figure 7 for an example).