Table 3: Electromagnetic Emissions 
Guidance and manufacturer’s declaration – electromagnetic emissions 
The Model 9010 Programmer System is intended for use in the electromagnetic environment specified 
below.  The customer or the user of the Model 9010 Programmer System should assure that it is used in 
such an environment. 
Emissions Test 
Compliance 
Electromagnetic  environment – guidance 
RF emissions 
CISPR 11 
Group 2 
The Model 9010 Programmer System must emit 
electromagnetic energy in order to perform its intended 
function.  Nearby electronic equipment may be affected. 
RF emissions 
CISPR 11 
Class B 
The Model 9010 Programmer System is suitable for use in all 
establishments, including domestic establishments and those 
directly connected to the public low-voltage power supply 
network that supplies buildings used for domestic purposes. 
Harmonic emissions 
IEC 61000-3-2 
Class A 
Voltage fluctuations / 
flicker emissions 
IEC 61000-3-3 
Complies