BAROSTIM NEO LEGACY REFERENCE GUIDE
18-3
Table 4: Electromagnetic Immunity
Guidance and manufacturer’s declaration – electromagnetic immunity
The Model 9010 Programmer System is intended for use in the electromagnetic environment specified below.
The customer or the user of the Model 9010 Programmer System should assure that it is used in such an
environment.
The customer or the user of the Model 9010 Programmer System should assure that it is used in such an
environment.
Immunity Test
IEC 60601 test level
Compliance level
Electromagnetic environment –
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 8 kV air
± 6 kV contact
± 8 kV air
± 8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30 %.
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
transient/burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/output
lines
supply lines
± 1 kV for input/output
lines
± 2 kV for power supply
lines
± 1 kV for input/output
lines
lines
± 1 kV for input/output
lines
Mains power quality should be that
of a typical commercial or hospital
environment.
of a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
IEC 61000-4-5
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
± 2 kV line(s) to earth
± 1 kV differential mode
± 2 kV common mode
± 2 kV common mode
Mains power quality should be that
of a typical commercial or hospital
environment.
of a typical commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5 % U
T
(>95 % dip in U
T
for
0,5 cycle)
40 % U
40 % U
T
(60 % dip in U
T
for 5
cycles)
70 % U
70 % U
T
(30 % dip in U
T
for 25
cycles)
<5 % U
<5 % U
T
(>95 % dip in U
T
for 5
s)
<5 % U
T
(>95 % dip in U
T
for 0,5
cycle)
40 % U
40 % U
T
(60 % dip in U
T
for 5
cycles)
70 % U
70 % U
T
(30 % dip in U
T
for 25
cycles)
<5 % U
<5 % U
T
(>95 % dip in U
T
for 5 s)
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the
Model 9010 Programmer System
requires continued operation
during power mains interruptions, it
is recommended that the Model
9010 Programmer System be
powered from an uninterruptible
power supply or a battery.
of a typical commercial or hospital
environment. If the user of the
Model 9010 Programmer System
requires continued operation
during power mains interruptions, it
is recommended that the Model
9010 Programmer System be
powered from an uninterruptible
power supply or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic of
a typical location in a typical
commercial or hospital
environment.
should be at levels characteristic of
a typical location in a typical
commercial or hospital
environment.
NOTE U
T
is the a.c. mains voltage prior to application of the test level.