CAS Corporation RW2601PRF Manual De Usuario

Descargar
Página de 19
 
(1) Normal Mode
RFErr
 Reason for error
     When RW-2601P can't communicate with RW-P in 5 seconds
     When RW-P has any problem.
    Treatment
     After check F12,F13, Turn off and then on.
 Err 06
  
 Reason
     Error in printer connection
    Trouble shooting
     Check the printer connector
     If there is no problem with printer connector, call your CAS dealer.
 Err 08
  
 Reason
     ZERO or TARE key does not work because weight is not stable.
    Trouble shooting
     When the weight is stable press the ZERO or TARE key.
 Err 09
  
 Reason
     The weight is beyond ZERO key range.
    Trouble shooting
     Press the ZERO key within ZERO key range.
 Err 13
  
 Reason
     The zero range deviates from the set range.
    Trouble shooting
     Confirm that there is nothing on the weighing platform.
     If there is nothing on the platform, do calibration in CAL mode.