Over
  
 Reason
     Weight is exceed maximum capacity of the scale.
    Trouble shooting
     Do not exceed the weight limit.
     
If the load cell is damaged, the load cell should be replaced.
 (2) CAL MODE
 Err 21
  
 Reason
     
 The resolution is set to be exceeded the limit 1/30,000.
    Trouble shooting
     
Lower the resolution.
     The resolution = Maximum weight/one division
     Modify the maximum weight in CAL1 or modify the division in CAL2 so that the resolution
should be below 1/30,000.
 Err 22
  
 Reason
    The weight for span calibration is set to be lower than 10 % of the maximum capacity
of the scale.
    Trouble shooting
     Set the weight for span calibration in CAL3 to be more than 10% of the
maximum capacity.
 Err 23
  
 Reason
     The weight for span calibration is set to be exceeded 100 % of the maximum capacity of the scale.
    Trouble shooting
     Set the weight for span calibration to be within the maximum capacity of  the scale in CAL 1.