7
Safety Instructions
Operating conditions
Caution!
The ICS may be operated only in areas used for 
medical purposes (in accordance with DIN VDE 
0107:1994). Do NOT operate the ICS 3000 in areas 
where there is a risk of explosion.
Note:
The ICS 3000 is designed to be operated and stored in 
an enclosed area.
Operate the ICS 3000 and its individual components 
only after placing it on a stable, level surface (e.g., a 
table).
Caution!
Connect only the BIOTRONIK power supply cable to 
the power supply; never use any other cables for the 
ICS 3000.
WARNING!
Never connect the ICS 3000 to the patient at the same 
time you are using electrosurgical instruments (such 
as an electrocautery). This might harm the patient 
and/or cause improper or unpredictable functioning of 
the device.
WARNING!
Under no circumstances should you attempt to 
change settings by selecting the parameter(s) several 
times in rapid succession. This could produce 
unintended results.
Caution!
The ICS 3000 has a touch screen for all input. Exercise 
care in configuring the settings, so that you do not 
activate an undesired function unintentionally.
WARNING!
Use the safe program function only under the direct 
supervision of a physician.
WARNING!
Keep an external defibrillator available when using 
the NIPS function.