Introduction
Chapter 1
5
DataRecorder 2 LED Indications
The following table describes the LED indicators and their status/color for each of the most 
common DataRecorder 2 events/status. 
LEDs
DataRecorder 2 Status
DataRecorder is ON but not initialized. DataRecorder does not capture capsule signals.
DataRecorder is initialized with patient data and ready to capture capsule signals.
DataRecorder shuts down if no capsule signals are received for more than 30, 60, or 90 
minutes, depending on the DataRecorder software version.
DataRecorder is exchanging status or data with RAPID or RAPID RT.
LED blinking rate varies according to the communication flow.
DataRecorder is capturing capsule signals.
Blinking rate = capsule frame rate.
DataRecorder has stopped capturing capsule signals for more than 5 seconds.
DataRecorder is detecting a capsule in sleep mode. 
Blinking rate = every five seconds (in any color).
DataRecorder is malfunctioning.
<20 seconds
DataRecorder is synchronizing with a capsule. This is normal functioning.
>20 seconds
DataRecorder detects capsule signal, but is not recording it. This is a malfunction.
Note
Check the SensorArray connection or have patient move to a different 
location.
Maximum Level
Battery Charge level
When charging, the Battery LEDs do not blink.
When DataRecorder is out of the Cradle, the Battery LEDS blink once 
every 5 seconds.
25%
Below 10%