PillCam Platform
6
Chapter 1
DataRecorder 2 Cradle 
The DataRecorder 2 Cradle is used to charge the DataRecorder 2 or to charge a spare battery 
externally. It is also used to discharge the battery before starting the recharge, when the Cradle 
detects that the battery needs refreshing (i.e., the battery gauge needs calibration). Thus 
occasionally, when inserted into the cradle, before charging starts, the Cradle may discharge first 
the battery and then start recharging.
The cradle also connects the DataRecorder 2 to the computer for performing patient check-in and 
creating a video.The green LED on the cradle indicates that the DataRecorder 2 is charged and 
ready for use.
The red LED, when lit continuously, indicates a defective battery.
The red LED, when blinking, indicates that there is a problem with the cradle.
Warning
Never connect the DataRecorder 2 to the SensorArray while the DataRecorder 
2 is in its cradle.
Cradle Connections
There are four connections on the back panel of the cradle. Only two of them are used with 
standard operation of the cradle: the power connector and the USB cable connection.
Note
When connecting more than one DataRecorder 2 to the computer, it is 
recommended to use a USB-powered hub.
The D-type connector and Auxiliary mains socket-outlet are for service use only.
Auxiliary mains socket
D-type connector
USB cable
Power connector