Technical Description
Chapter 5
49
Guidance and manufacturer’s declaration - electronic emissions
The PillCam capsules are intended for use in the electromagnetic environment specified below. The 
customer or the user of the PillCam capsules should assure that it is used in such an environment.
Immunity test
IEC 60601 test 
level
Compliance 
level
Electromagnetic 
environment - guidance
Electrostatic discharge 
(ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or 
ceramic tile. If floors are covered with 
synthetic material, the relative 
humidity should be at least 30 %.
Electrical fast transient 
/ burst
IEC 61000-4-4
±2 kV for power supply 
lines
±1 kV for input/output 
lines
Not applicable
Not applicable
Surge,
IEC 61000-4-5
±1 kV line(s) to line(s)
±2 kV line(s) to earth
Not applicable
Not applicable
Voltage dips, short 
interruptions and 
voltage variations on 
power supply input 
lines
IEC 61000-4-11
<5 % U
T
 
(>95 % dip in U
T
)
for 0.5 cycle
40 % U
T
 
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
<5 % U
T
 
(>95 % dip in U
T
)
for 5 sec
Not applicable
Not applicable
Power frequency (50/
60 Hz) magnetic field, 
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields 
should be at levels characteristic of a 
typical location in a typical commercial 
or hospital environment.
NOTE: U
T
   is the AC mains voltage prior to application of the test level.