Technical Description
Chapter 5
51
PillCam COLON 2 Capsules
Recommended separation distances between portable and mobile RF
communications equipment and the PillCam capsules
The PillCam capsules are intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the PillCam capsules can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the PillCam capsules as recommended below, according to
the maximum output power of the communications equipment.
disturbances are controlled. The customer or the user of the PillCam capsules can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the PillCam capsules as recommended below, according to
the maximum output power of the communications equipment.
Rated maximum
output power of
transmitter [W]
Separation distance according to frequency of transmitter [m]
150 kHz to 80 MHz
d = 1.2 P
80 MHz to 800 MHz
d = 1.2 P
800 MHz to 2,5 GHz
d = 2.3 P
0.01
Not applicable
0.12
0.23
0.1
Not applicable
0.38
0.73
1
Not applicable
1.2
2.3
10
Not applicable
3.8
7.3
100
Not applicable
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d in meters (m) can be determined using the equation applicable to the frequency of the transmitter, where
P is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
d in meters (m) can be determined using the equation applicable to the frequency of the transmitter, where
P is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Guidance and manufacturer’s declaration - electromagnetic emissions
The PillCam COLON 2 capsules are intended for use in the electromagnetic environment specified below.
The customer or the user of the PillCam COLON 2 capsules capsule should assure that it is used in such
an environment.
The customer or the user of the PillCam COLON 2 capsules capsule should assure that it is used in such
an environment.
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
CISPR 11
Group 1
The PillCam COLON 2 capsules use RF energy only
for its internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment
for its internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment
RF emissions
CISPR 11
CISPR 11
Class B
The PillCam capsules are suitable for use in all
establishments including domestic establishments and
those directly connected to the public low-voltage
power supply network that supplies buildings used for
domestic purposes.
establishments including domestic establishments and
those directly connected to the public low-voltage
power supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC 61000-3-2
IEC 61000-3-2
Not applicable
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
flicker emissions
IEC 61000-3-3
Not applicable